Article Text

Download PDFPDF
Case report
Mechanical revascularization of acute iatrogenic anterior cerebral artery occlusions: use of a new coaxial dual-lumen balloon catheter results in rapid access and flow restoration

Abstract

Acute iatrogenic occlusion of cerebral vessels is a risk during the performance of neuroendovascular procedures. Most commonly this is a result of thrombus formation within the vessel, thromboembolism or an acute vasospastic response. There are several options for flow restoration including pharmacological, mechanical and microsurgical. Mechanical flow restoration, usually by thrombectomy, is typically performed in large cerebral vessels. Current mechanical thrombectomy devices have limited utility in smaller vessels that are more difficult to access, such as the anterior cerebral artery (ACA). We present two cases where successful flow restoration of an intraoperative acutely occluded or thrombotic ACA was rapidly achieved by balloon dilation using a new coaxial dual-lumen balloon catheter. The time elapsed from the decision to use the balloon catheter until revascularization for each patient was 5 min and 46 s (case 1) and 10 min and 25 s (case 2).

  • Thrombectomy
  • Angioplasty
  • Balloon
  • Blood Flow
  • Complication

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.