Distal access using hyperflexible atraumatic distal tip with optimized proximal stability of the Benchmark intracranial guide catheter for the treatment of cerebral vascular diseases: a technical note
  1. Arun Chandran1,
  2. Mani Puthuran1,
  3. Paul R Eldridge2,
  4. Hans C Nahser1
  1. 1Neuroradiology Department, The Walton Centre NHS Foundation Trust, Liverpool, UK
  2. 2Neurosurgery Department, The Walton Centre NHS Foundation Trust, Liverpool, UK
  1. Correspondence to Dr Arun Chandran, Neuroradiology Department, The Walton Centre, Lower Lane, Fazakerley, Liverpool L9 7 LJ, UK; Arun.Chandran{at}thewaltoncentre.nhs.uk

Abstract

Background A stable guide catheter position within the intracranial vasculature is critical for safe, successful endovascular treatment.

Objective To present ourinitial experience with the 0.071 inch inner diameter Benchmark guide catheter used in the treatment of intracranial cerebrovascular pathologies, demonstrating its safety and efficacy.

Methods We retrospectively reviewed use of the Benchmark guide catheter from September through December 2014 in the management of various neuroendovascular intracranial pathologies. Clinical performance and complication rates were evaluated, with particular consideration of vessel tortuosity. A total of 62 Benchmarks were used, 47 in the anterior circulation, 10 in the posterior circulation, 4 in the external carotid, and 1 in the venous sinus. The five cases with access to the external carotid and venous sinus were excluded.

Results The Benchmark was able to cross at least one 90° turn in 49 (86%) of the 57 patients. Reversal of the catheter was seen in 15% of 47 anterior circulation cases (4 at one 90° turn; 3 at two 90° turns). We report no complications of dissection or thromboembolic events. All guide catheter positions were safely achieved over a 0.035 Terumo stiff glidewire without need for an inner smaller lumen guide catheter for navigation.

Conclusions Benchmark is a new guide catheter, with an ideal combination of both hyperflexible, atraumatic distal tip and optimized proximal shaft support to provide stable 6F primary access for a successful neurointerventional procedure. Benchmark can be easily, safely, and consistently positioned in a desired location within intracranial arteries providing a stable position for intervention and adequate angiography.

  • Catheter
  • Brain
  • Intervention
  • Artery
  • Device

This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/