Article Text

Download PDFPDF
Original research
Initial and mid-term results from 108 consecutive patients with cerebral aneurysms treated with the WEB device
  1. Christin Clajus1,
  2. Christoph Strasilla1,
  3. Tom Fiebig2,
  4. Vojtech Sychra1,
  5. David Fiorella3,
  6. Joachim Klisch1
  1. 1Department of Diagnostic and Interventional Radiology and Neuroradiology, Helios General Hospital Erfurt, Erfurt, Germany
  2. 2Department of Diagnostic and Interventional Radiology, Helios General Hospital Meiningen, Meiningen, Germany
  3. 3Department of Neurosurgery, State University of New York at Stony Brook, Stony Brook, New York, USA
  1. Correspondence to Dr Christin Clajus, Department of Diagnostic and Interventional Radiology and Neuroradiology, Helios General Hospital Erfurt, Nordhäuser Str. 74, Erfurt 99089, Germany; christin.clajus{at}googlemail.com

Abstract

Introduction The Woven EndoBridge (WEB) is a novel device for the treatment of wide-necked intracranial bifurcation aneurysms. The present series demonstrates our ‘real-world experience’ in the use of all iterations of WEB devices (available in Europe) in ruptured and unruptured aneurysms.

Methods We analyzed our all-inclusive cerebrovascular database for patients treated with the WEB device between October 2010 and May 2015. Anatomic and clinical results are reported for all patients.

Results One hundred and eight patients with 114 intracranial aneurysms were included in the series. Forty-seven aneurysms (41.2%) were ruptured. Eighty-six patients received angiographic and clinical follow-up after a mean of 13.4 months. One hundred and ten of 114 WEB devices (96.5%) were deployed successfully. Thromboembolic complications occurred in 11 of 110 interventions (10.0%), with a new permanent deficit in one patient. Re-rupture after WEB treatment was detected in two aneurysms (4.3%), which had both initially presented with subarachnoid hemorrhage. Angiographic follow-up revealed adequate occlusion in 68 of 90 aneurysms (75.6%). Fifteen aneurysms required retreatment.

Conclusions This series confirms a high level of safety and efficacy of the WEB device for the treatment of wide-necked intracranial aneurysms.

  • Aneurysm
  • Device
  • Flow Diverter
  • Hemorrhage

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.