Responses

Download PDFPDF
Case report
Republished: Stentriever thrombectomy with distal protection device for carotid free floating thrombus: a technical case report
Compose Response

Plain text

  • No HTML tags allowed.
  • Web page addresses and e-mail addresses turn into links automatically.
  • Lines and paragraphs break automatically.
Author Information
First or given name, e.g. 'Peter'.
Your last, or family, name, e.g. 'MacMoody'.
Your email address, e.g. higgs-boson@gmail.com
Your role and/or occupation, e.g. 'Orthopedic Surgeon'.
Your organization or institution (if applicable), e.g. 'Royal Free Hospital'.
Statement of Competing Interests

PLEASE NOTE:

  • A rapid response is a moderated but not peer reviewed online response to a published article in a BMJ journal; it will not receive a DOI and will not be indexed unless it is also republished as a Letter, Correspondence or as other content. Find out more about rapid responses.
  • We intend to post all responses which are approved by the Editor, within 14 days (BMJ Journals) or 24 hours (The BMJ), however timeframes cannot be guaranteed. Responses must comply with our requirements and should contribute substantially to the topic, but it is at our absolute discretion whether we publish a response, and we reserve the right to edit or remove responses before and after publication and also republish some or all in other BMJ publications, including third party local editions in other countries and languages
  • Our requirements are stated in our rapid response terms and conditions and must be read. These include ensuring that: i) you do not include any illustrative content including tables and graphs, ii) you do not include any information that includes specifics about any patients,iii) you do not include any original data, unless it has already been published in a peer reviewed journal and you have included a reference, iv) your response is lawful, not defamatory, original and accurate, v) you declare any competing interests, vi) you understand that your name and other personal details set out in our rapid response terms and conditions will be published with any responses we publish and vii) you understand that once a response is published, we may continue to publish your response and/or edit or remove it in the future.
  • By submitting this rapid response you are agreeing to our terms and conditions for rapid responses and understand that your personal data will be processed in accordance with those terms and our privacy notice.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.

Vertical Tabs

Other responses

Jump to comment:

  • Published on:
    Aspiration with distal filter protection: an effective management for carotid floating thrombus
    • Biplab Das, Neurointervention Medanta The Medicity, Gurgaon, India
    • Other Contributors:
      • Gaurav Goel, Neurointervention
      • Anshu Mahajan, Neurointervention
      • Arun Garg, Neurology
      • Harsh Sapra, Neuroanaesthesia

    To the editor,

    Giragani S et al. (published online 25 January 2017) described a remarkable case of recurrent transient ischemic attacks (TIA) due to carotid free floating thrombus. They successfully used stentriever with distal filter protection in retrieving the thrombus.

    Here we share a similar case of TIA with right common carotid artery (CCA) floating thrombus that was effectively managed with distal filter protection and aspiration. A 48-year-male with recurrent ischemic symptoms detected to have right CCA long segment floating thrombus (approximately 4.5 cm) extending upto right proximal external carotid artery.* Under general anaesthesia through right femoral route long sheath guiding catheter (Neuron Max 6F088; Penumbra, Inc. Alameda, USA) was placed in right proximal CCA. After parking the filter device (Spider FX 6mm; eV3, Plymouth, Minnesota, USA) at distal cervical segment, thrombus was aspirated using penumbra system (5MAX ACE, 132 cm; Penumbra, Inc. Alameda, USA).* Final check angiography showed 80 % reduction in clot burden .*

    Placing the filter protection device in the distal cervical segment does not protect thrombus migration to ECA, although it primarily prevent intracranial shower. In our index case thrombus fragment migrated to ECA, although it did not cause any neurological deficit. These cases highlight a novel technique to treat free floating thrombus.

    *Representative image available.

    Conflict of Interest:
    None declared.