RT Journal Article SR Electronic T1 Flow diverter implantation in a rat model of sidewall aneurysm: a feasibility study JF Journal of NeuroInterventional Surgery JO J NeuroIntervent Surg FD BMJ Publishing Group Ltd. SP 88 OP 92 DO 10.1136/neurintsurg-2016-012878 VO 10 IS 1 A1 René Aquarius A1 Debby Smits A1 Matthew J Gounis A1 William P J Leenders A1 Joost de Vries YR 2018 UL http://jnis.bmj.com/content/10/1/88.abstract AB Background More challenging animal models are needed to elucidate the efficacy of flow diverter (FD) designs and the mechanisms behind observed complications. The purpose of this study is to demonstrate the feasibility of implanting a FD in a sidewall aneurysm rat model.Methods An end-to-side anastomosis was created in the abdominal aorta of 36 rats using a decellularized donor pouch. A FD was subsequently implanted.Results After up to 3 months of follow-up, we observed that rats displayed normal growth and behavior. Mortality within the groups was low (2 rats, 5.6%). All aneurysms thrombosed after FD implantation and showed progressive soft tissue replacement of the thrombus during follow-up. The abdominal aortas remained patent.Conclusions This model can be used to test the effects of FDs in future studies.