TableĀ 1

Technical problems encountered during procedure and causes of procedure failure

Patient sequenceStenosis locationStenosis degreeTechnical problem encountered during the procedureCause of procedure failure
4ICA and MCA50%Difficult stent tracking and delivery, fracture core of stent delivery system
5ICA60%Nose cone bending caused stent sheath kinking and stretching during stent deployment
7ICA60%Stretching of stent sheath during stent deployment
8BA75%Arterial rupture after angioplasty
15ICA and MCA50%TIA during manipulation of balloon catheter across stenosisTIA
27ICA90%Nose cone trapped between guidewire and stent tip after stent deployment
42ICA75%Difficult stent tracking and delivery, fracture core of stent delivery system, TIATIA
44MCA83%TIA
47ICA and MCA96%Failure passage of balloon across stenosisFailure passage of balloon
48MCA90%Kinking of balloon catheter after guidewire removal
56MCA70%Transient blindness and severe headache during stent system advancement
59MCA87%Transient blindness during stent system advancementTIA
66ICA75%Perforation of M2 perforators by guidewire during angioplasty
  • BA, basilar artery; ICA, internal carotid artery; MCA, middle cerebral artery; TIA, transient ischemic attack.