Patient sequence | Stenosis location | Stenosis degree | Technical problem encountered during the procedure | Cause of procedure failure |
---|---|---|---|---|
4 | ICA and MCA | 50% | Difficult stent tracking and delivery, fracture core of stent delivery system | |
5 | ICA | 60% | Nose cone bending caused stent sheath kinking and stretching during stent deployment | |
7 | ICA | 60% | Stretching of stent sheath during stent deployment | |
8 | BA | 75% | Arterial rupture after angioplasty | |
15 | ICA and MCA | 50% | TIA during manipulation of balloon catheter across stenosis | TIA |
27 | ICA | 90% | Nose cone trapped between guidewire and stent tip after stent deployment | |
42 | ICA | 75% | Difficult stent tracking and delivery, fracture core of stent delivery system, TIA | TIA |
44 | MCA | 83% | TIA | |
47 | ICA and MCA | 96% | Failure passage of balloon across stenosis | Failure passage of balloon |
48 | MCA | 90% | Kinking of balloon catheter after guidewire removal | |
56 | MCA | 70% | Transient blindness and severe headache during stent system advancement | |
59 | MCA | 87% | Transient blindness during stent system advancement | TIA |
66 | ICA | 75% | Perforation of M2 perforators by guidewire during angioplasty |