Table 2

Result of Flat plate tests and Pull up traction tests

SizeFlat plate tests
Radial force density
N/mm
Pull up traction tests
Radial pressure Pa (N/m2) (1.5 mm tube)
Pull up traction tests
Radial pressure Pa (N/m2) (3.5 mm tube)
Trevo PV
 4–200.0148092050* (p=0.050)
 3–200.0060030060* (p=0.030)
Eric
 3–200.01100220
 4–240.018501540340* (p=0.008)
 6–440.01304770400* (p=0.010)
Embotrap
 5–210.006421430770* (p=0.014)
Separator 3D
 4.5–260.007911360400* (p=0.006)
Revive
 4.5–220.012691360850* (p=0.021)
Mindframe
 3–230.004511250330* (p=0.034)
Solitaire FR
 4–200.004481110530* (p=0.014)
 6–300.003511060580* (p=0.020)
Preset
 4–200.005211090730* (p=0.018)
 6–300.00368630600* (p=0.25)
Preset LT
 3–200.003201060160* (p=0.03)
 4–200.00370460480* (p=0.77)
Catch
 3–150.00350840220* (p=0.049)
 4–200.004441810360* (p=0.010)
 6–300.003681870900* (p=0.030)
  • *p Value for the comparison between force exerted in tubes of 1.5 and 3.5 mm inner diameter during Pull up traction tests. Note results obtained with Preset 6–30 and Preset LT 4–20 for which there is not significant shift of the radial pressure when retrieved in tubes of different diameter; p=0.25 and p=0.77, respectively.