Abstract O-029 Table 2 Efficacy of ADAPT Technique with Evolving Aspiration Catheter Technology

First aspiration catheter used:Mean time puncture to reperfusion (min):TICI 2b/3 after 1 ADAPT pass:TICI 2b/3 with ADAPT only:Mean total number of ADAPT passes:Solumbra rescue rate:
3 Max: (n=2) 55.5 0 0 2.5 100
ACE 60: (n=45) 48.8 31% 64% 2.4 29%
ACE 64: (n=34) 30.2 38% 88% 2.5 3%
ACE 68: (n=16) 29.3 50% 100% 2.6 0
NeuronMax: (n=3) 22.7 100% 100% 1 0
p-value: 0.002 (ACE 60 vs ACE 64/68) 0.09 <0.001 0.78 (ACE 60 vs ACE 64/68) <0.001