Comparison of devices used for stent-assisted coiling of intracranial aneurysms

PLoS One. 2011;6(9):e24875. doi: 10.1371/journal.pone.0024875. Epub 2011 Sep 22.

Abstract

Introduction: Two self-expandable stents, the Neuroform and the Enterprise stent, are widely used for stent-assisted coiling (SAC) of complex shaped intracranial aneurysms. However, comparative knowledge about technical feasibility, peri- and post-procedural morbidity and mortality, packing densities as well as follow-up data is limited.

Material and methods: We conducted a retrospective study to investigate differences in aneurysms stented with the Enterprise or Neuroform stents. Angiographic follow-up (mean 19.42 months) was available in 72.6% (61/84) of aneurysms treated with stent-assisted coiling. We further sought to compare stent-assisted coiling to a matched patient population with aneurysms treated by conventional coil embolization.

Results: The stenting success rate of the Enterprise was higher compared to the Neuroform stent (46/48 and 42/51, respectively). In 5 of 9 cases in which the Neuroform stent was not navigable to the landing zone, we successfully deployed an Enterprise stent instead. Eventually, 42 aneurysms were coiled after stenting in each group. We observed no significant differences in peri-procedural complication rate, post-procedural hospital stay, packing density, recurrence rate or number of in-stent stenosis. Strikingly, 36.1% of followed aneurysms in the SAC group showed progressive occlusion on angiographic follow-up imaging. The packing density was significantly higher in aneurysms treated by SAC as compared to conventionally coiled aneurysms, while recanalization rate was significantly lower in the SAC group.

Conclusion: The procedural success rate is higher using the Enterprise, but otherwise both stents exhibited similar characteristics. Lower recurrence frequency and complication rates comparable to conventional coil embolization emphasize the importance of stent-assisted coiling in the treatment of complex aneurysms. Progressive occlusion on angiographic follow-up was a distinct and frequent observation in the SAC group and may in part be due to flow diversion.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Cerebral Angiography
  • Embolization, Therapeutic / instrumentation
  • Embolization, Therapeutic / methods*
  • Feasibility Studies
  • Female
  • Follow-Up Studies
  • Humans
  • Intracranial Aneurysm / diagnostic imaging
  • Intracranial Aneurysm / therapy*
  • Male
  • Middle Aged
  • Reproducibility of Results
  • Retrospective Studies
  • Stents / classification
  • Stents / standards*
  • Treatment Outcome