Catheter exchange technique for the Wingspan stent: technical note
  1. W S Lesley,
  2. R Rangaswamy
  1. Texas A&M Health Sciences Center College of Medicine, Scott & White Clinic, Temple, Texas, USA
  1. Correspondence to Walter S Lesley, Section of NeuroInterventional Surgery, Scott & White Clinic, 2401 South 31st Street, Temple, Texas 76508, USA; wlesley{at}swmail.sw.org

Abstract

Background and purpose The only Food and Drug Administration-approved implant for managing intracranial stenosis is the Wingspan stent. Modification of the standard Wingspan stent delivery procedure has not been previously reported.

Material and methods Retrospective, single-patient, technical report.

Results Following uneventful balloon angioplasty, a symptomatic intracranial vertebral artery stenosis was inaccessible to Wingspan stent delivery using standard procedural protocol because of poor proximal guide catheter purchase. After placing a Renegade Hi-Flo microcatheter across the stenosis, the Wingspan stent was transferred to this catheter and advanced to the lesion using a Neuroform stabilizer. Unsheathing of the stent yielded uncomplicated, precise placement of the stent with a good clinical result.

Conclusion The Wingspan stent transfer technique can be used successfully in the management of intracranial stenotic disease.

  • Catheter
  • Stent
  • Technique

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Competing interests None.

Linked Articles

  • Correction
    BMJ Publishing Group Ltd.BMA House, Tavistock Square, London, WC1H 9JR
  • Correction
    BMJ Publishing Group Ltd.BMA House, Tavistock Square, London, WC1H 9JR