Article Text
Abstract
Background and purpose The only Food and Drug Administration-approved implant for managing intracranial stenosis is the Wingspan stent. Modification of the standard Wingspan stent delivery procedure has not been previously reported.
Material and methods Retrospective, single-patient, technical report.
Results Following uneventful balloon angioplasty, a symptomatic intracranial vertebral artery stenosis was inaccessible to Wingspan stent delivery using standard procedural protocol because of poor proximal guide catheter purchase. After placing a Renegade Hi-Flo microcatheter across the stenosis, the Wingspan stent was transferred to this catheter and advanced to the lesion using a Neuroform stabilizer. Unsheathing of the stent yielded uncomplicated, precise placement of the stent with a good clinical result.
Conclusion The Wingspan stent transfer technique can be used successfully in the management of intracranial stenotic disease.
- Catheter
- Stent
- Technique
Statistics from Altmetric.com
Footnotes
Competing interests None.
Linked Articles
- Correction
- Correction