In vitro measurement of the permeability of endovascular coils deployed in cerebral aneurysms
  1. Chander Sadasivan,
  2. Erica Swartwout,
  3. Ari D Kappel,
  4. Henry H Woo,
  5. David J Fiorella,
  6. Barry B Lieber
  1. Department of Neurological Surgery, Stony Brook University, Stony Brook, New York, USA
  1. Correspondence to Chander Sadasivan, Department of Neurosurgery, Stony Brook University, Stony Brook, NY 11794, USA; csadasivan{at}sbumed.org

Abstract

Background and purpose Aneurysm recurrence is the primary limitation of endovascular coiling treatment for cerebral aneurysms. Coiling is currently quantified by a volumetric porosity measure called packing density (pd). Blood flow through a coil mass depends on the permeability of the coil mass, and not just its pd. The permeability of coil masses has not yet been quantified. Here we measure coil permeability with a traditional falling-head permeameter modified to incorporate idealized aneurysms.

Methods Silicone replicas of idealized aneurysms were manufactured with three different aneurysm diameters (4, 5, and 8 mm). Four different coil types (Codman Trufill Orbit, Covidien Axium, Microvention Microplex 10, and Penumbra 400) were deployed into the aneurysms with a target pd of 35%. Coiled replicas were installed on a falling-head permeameter setup and the time taken for a column of fluid above the aneurysm to drop a certain height was recorded. Permeability of the samples was calculated based on a simple modification of the traditional permeameter equation to incorporate a spherical aneurysm.

Results The targeted 35% pd was achieved for all samples (35%±1%, P=0.91). Coil permeabilities were significantly different from each other (P<0.001) at constant pd. Microplex 10 coils had the lowest permeability of all coil types. Data suggest a trend of increasing permeability with thicker coil wire diameter (not statistically significant).

Conclusions A simple in vitro setup was developed to measure the permeabilities of coil masses based on traditional permeametry. Coil permeability should be considered when evaluating the hemodynamic efficacy of coiling instead of just packing density. Coils made of thicker wires may be more permeable, and thus less effective, than coils made from thinner wires. Whether aneurysm recurrence is affected by coil wire diameter or permeability needs to be confirmed with clinical trials.

  • aneurysm
  • blood flow
  • coil
  • device

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Contributors Author CS contributed to study conception, data collection, and interpretation as well as manuscript preparation. Author ES contributed to the study design, data collection, and analysis. Author ADK contributed to the data collection and analysis. Authors DJF and HHW contributed to the overall study conception and design. Author BBL contributed to important intellectual content of the study. All authors agree to be accountable for all aspects of the work.

  • Competing interests Author HHW receives royalties from Codman. Author DJF has consulted for Medtronic, has received unrelated research support from, as well as served as consultant for, Penumbra and Microvention, and receives royalties from Codman.

  • Provenance and peer review Not commissioned; externally peer reviewed.