Medina embolization device for the treatment of intracranial aneurysms: 18 months’ angiographic results

Abstract

Background and purpose The Medina embolization device (MED) is a new flow disruption device combining the design of a detachable coil with an intrasaccular flow disrupter. Safety and short-term angiographic effectiveness of this device have recently been reported. However, long-term angiographic results are lacking. We report herein the 18 months’ angiographic outcome in patients treated for a wide-neck intracranial aneurysm with the MED.

Materials and methods Nineteen patients (17 female, mean age 50 years) with 20 wide-neck intracranial aneurysms (six ruptured; 14 unruptured) were treated by the MED between January 2015 and June 2016. Procedure-related complications were systematically recorded; discharge and 6–9 months' follow-up modified Rankin Scale scores were assessed. Angiographic mid-term and long-term follow-up were performed with a mean delay of 6.4±1.5 months (n=16 aneurysms) and 17.7±4.2 months (n=15 aneurysms), respectively. Occlusion rates were evaluated after the procedure and at the mid-term and long-term follow-up using the Roy-Raymond scale.

Results Embolization with the MED was feasible in all except two cases (2/20, 10%). One per-procedural perforation was recorded (1/20, 5%) and one MED deployment failed because of the aneurysm’s shape (1/20, 5%). Three cases of thromboembolic complications were observed (3/20, 15%). Only one thromboembolic complication was responsible for clinical sequelae. Grade A occlusion rate was 61% (11/18) after the procedure, 75% at 6 months' follow-up (12/16), and 80% (12/15) at long-term follow-up. Two cases (2/18, 11%) of recanalization at mid-term were documented angiographically. No recanalization occurred between the mid-term and long-term follow-up.

Conclusion MED is a hybrid embolization device, combining properties of a conventional coil with those of an intrasaccular flow disrupter. Our series focusing on long-term angiographic follow-up shows a satisfactory long-term occlusion rate. Larger series with longer angiographic follow-up times are warranted to confirm these preliminary results.

  • aneurysm
  • device
  • neck

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.