Article Text

Download PDFPDF

P-021 COMFORT- Colombian multicenter flow-diverter observational reconstruction trial. local experience in the endovascular treatment of intracranial aneurysms with FRED stent
Free
  1. B Pabon1,
  2. C Diaz1,
  3. M Fonseca2,
  4. N Lobelo3,
  5. J Holguin4,
  6. J Mejia1,
  7. O Vargas1,
  8. M Patiño1,
  9. N Serna1
  1. 1Neuroendovascular Surgery, Angioteam – Clinica del Norte, Medellin, Colombia
  2. 2Neuroendovascular Surgery, unipamplona. medinorte, Cucuta, Colombia
  3. 3Neuroendovascular Surgery, Clinica Colombia, bogota, Colombia
  4. 4Neuroendovascular Surgery, Clinica Imbanaco, Cali, Colombia

Abstract

Purpose Flow-diverting stents, such as the Flow Re-direction Endoluminal Device (FRED; MicroVention, Tustin, California, USA), have emerged as a novel means of treating complex intracranial aneurysms. This observational analysis of the initial Colombian experience provides insight into patient selection, technical challenges, clinical and radiographic outcomes, and complication rates after the use of FRED device for intracranial aneurysms.

Materials and methods Cases were compiled from 5 Colombian centers between March 2014 and April 2017. We performed a multicenter study evaluating patients with intracranial aneurysms treated with FRED. Technical success, morbidity and mortality were registered. Regarding safety, a well-defined variables included: spontaneous rupture of the FRED-treated aneurysm; spontaneous nonaneurysmal or distal intracranial hemorrhage (ICH); thrombo- embolic stroke; parent artery stenosis, and permanent cranial neuropathy.

Results During the defined study period, 252 patients with 268 aneurysms treated with FRED were included in this registry. Technical success stent deployment observed in all cases with exception of two cases where the operators decided to pull back the device and treated the aneurysms with other strategy. The mean aneurysm size was 12.4 ± 5 mm, and the median angiographic follow-up was 11.5 months. 81 aneurysms (30.2%) were small, 132 (49.2%) were large and 55 (20.5%) were giant. The median clinical follow-up time was 16 months. The neurological morbidity rate was 7.5% (19/252), and the neurological mortality rate was 2.3% (6/252). The combined neurological morbidity/mortality rate was 9.8%. The most common adverse events were ischemic stroke (5.9%, 15/252) and spontaneous ICH (1.9%, 5/252). The complete occlusion rate at the last follow-up was 87.6% (92/105).

Conclusion In the largest regional series on FRED for intracranial aneurysms to date, data suggest that treatment with the Flow Re-direction Endoluminal Device is safe and efficacious, with complication rates comparable with others FD available. Our local results are promising but larger series with long-term follow-up are required to determine its superiority.

Disclosures B. Pabon: 2; C; Microvention, Medtronic, Stryker. C. Diaz: None. M. Fonseca: None. N. Lobelo: None. J. Holguin: None. J. Mejia: None. O. Vargas: None. M. Patiño: None. N. Serna: None.

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.