Article Text
Abstract
Background Stent-assisted coiling of wide-necked intracranial aneurysms (IAs) using the Neuroform Atlas Stent System (Atlas) has shown promising results.
Objective To present the primary efficacy and safety results of the ATLAS Investigational Device Exemption (IDE) trial in a cohort of patients with posterior circulation IAs.
Methods The ATLAS trial is a prospective, multicenter, single-arm, open-label study of unruptured, wide-necked, IAs treated with the Atlas stent and adjunctive coiling. This study reports the results of patients with posterior circulation IAs. The primary efficacy endpoint was complete aneurysm occlusion (Raymond-Roy (RR) class I) on 12-month angiography, in the absence of re-treatment or parent artery stenosis >50%. The primary safety endpoint was any major ipsilateral stroke or neurological death within 12 months. Adjudication of the primary endpoints was performed by an imaging core laboratory and a Clinical Events Committee.
Results The ATLAS trial enrolled and treated 116 patients at 25 medical centers with unruptured, wide-necked, posterior circulation IAs (mean age 60.2±10.5 years, 81.0% (94/116) female). Stents were placed in all patients with 100% technical success rate. A total of 95/116 (81.9%) patients had complete angiographic follow-up at 12 months, of whom 81 (85.3%) had complete aneurysm occlusion (RR class I). The primary effectiveness outcome was achieved in 76.7% (95% CI 67.0% to 86.5%) of patients. Overall, major ipsilateral stroke and secondary persistent neurological deficit occurred in 4.3% (5/116) and 1.7% (2/116) of patients, respectively.
Conclusions In the ATLAS IDE posterior circulation cohort, the Neuroform Atlas Stent System with adjunctive coiling demonstrated high rates of technical and safety performance.
Trial registration number https://clinicaltrials.gov/ct2/show/NCT02340585.
- aneurysm
- coil
- stent
Data availability statement
Data are available upon reasonable request. For inquiries, please contact: Brian T. Jankowitz, Associate Professor Neurosurgery Cooper University Hospital Camden, NJ 08103. Email: jankowitz-brian@CooperHealth.edu
This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/.
Statistics from Altmetric.com
Introduction
Wide-necked intracranial aneurysms (IAs) (neck ≥4 mm, dome-to-neck ratio <2) constitute at least 40% of all IAs,1–3 and are difficult to treat with endovascular coiling alone given the propensity for coils to herniate into the parent artery.4 5 Stent-assisted coiling (SAC) is a well-established endovascular treatment,6–9 which reconstructs the aneurysm neck, prevents coil herniation into the parent artery, and theoretically expedites aneurysm healing by creating a scaffold for endothelial coverage.5 10
The Neuroform stent system (Stryker Neurovascular, Fremont, California, USA) was first approved by the United States Food and Drug Administration (FDA) in 2002. Since then, newer stent iterations, such as Neuroform EX and EZ3, have been approved. The Neuroform Atlas Stent System (Atlas) is the newest generation.9 11–13 The device was designed to scaffold the aneurysm neck and support the placement of detachable, intrasaccular coils. Significant design advances include improved trackability and a smaller cell size, which provides better coil retention within the aneurysm. The Atlas stent design also allows for its delivery via a lower profile, 0.0165 inch (internal diameter) microcatheter as compared with the 0.027 inch microcatheter required for the original Neuroform stent. Finally, the new hybrid cell design, with closed cells at the proximal end, improves recrossing and enhances stability within the vessel, while the classic central open-cell design provides excellent wall apposition, conformability, and flexibility.
In a previous prospective Neuroform Atlas investigational device exemption (IDE) study of 182 patients with anterior circulation IAs, SAC via the Atlas stent demonstrated high rates of complete occlusion at the 12-month angiographic follow-up,14 15 as well as promising safety profiles in the anterior cerebral circulation.11 In this study, we present the primary efficacy and safety results of the ATLAS IDE trial in a cohort of patients with posterior circulation IAs.
Methods
Study design
The ATLAS IDE trial is a prospective, multicenter, open-label, non-randomized, two-cohort, single-arm study that enrolled patients with wide-necked intracranial saccular aneurysms to be treated with SAC using the Atlas stent at 25 medical centers in the USA. Safety and efficacy endpoints were evaluated in a modified intention-to-treat cohort of patients who signed the informed consent form and in whom the investigational device entered the body. An imaging core laboratory and an independent Clinical Events Committee (CEC) adjudicated the primary efficacy and safety endpoints, respectively, to ensure consistency and accuracy of the data and minimize bias.
The institutional review board at each enrollment center approved the study protocol. Each patient completed a written informed consent prior to participation in the trial. All data were entered into a Health Insurance Portability and Accountability Act (HIPAA)-compliant electronic data capture system and monitored by the sponsor and a contract research organization. The data used to support the conclusions of this trial will be furnished, on reasonable requests, by the corresponding author.
Patient enrollment
Enrollment for patients with posterior circulation IAs (including vertebral, basilar, and posterior cerebral arteries) took place between June 2015 and December 2017. Clinical investigators and designated research staff at each center managed patient identification, recruitment, and enrollment. Patients were considered enrolled in the study once the site investigator determined that they met all trial inclusion/exclusion criteria and provided the signed informed consent form. The full study enrollment criteria are provided in online supplemental table 1). Briefly, patients were included if they were 18–80 years old and had a documented, wide-necked (neck ≥4 mm or dome-to-neck ratio <2), intracranial, saccular aneurysm arising from a parent vessel with a diameter of 2.0–4.5 mm. The following criteria were used to determine study exclusion: multiple untreated IAs requiring treatment, acute target aneurysm rupture <14 days prior to study treatment, modified Rankin Scale (mRS) score of ≥4 or Hunt and Hess scale score ≥3, intracranial mass or cerebral vascular malformation, a target aneurysm in the anterior circulation proximal to the superior hypophyseal internal carotid artery, previous treatment with SAC embolization, a known absolute contraindication to angiography or antiplatelet therapy, Moyamoya disease, or underlying parent artery atherosclerosis.
Supplemental material
Procedure description
The procedure was previously described in detail in the primary results of the ATLAS humanitarian device exemption study.16 In brief, all patients undergoing treatment were premedicated with dual antiplatelet therapy (aspirin and clopidogrel) for at least 5 days. Platelet reactivity testing was not mandated. All procedures were performed under general anesthesia, and anticoagulation was managed according to each study site standard of care with a recommended activated clotting time of 250–300 s during the procedure. Atlas was deployed using an Excelsior SL-10 or XT-17 (Stryker Neurovascular, Fremont, California, USA) microcatheter via a transfemoral, radial, or brachial percutaneous approach. Dual antiplatelet therapy was maintained for at least 3 months following stent implantation.
Follow-up evaluation
After the implant procedure, all treated patients had follow-up evaluations within 72 hours of the procedure, prior to hospital discharge, and at 2, 6, and 12 months. Data were collected to assess primary and secondary endpoints. Data consisted of neurological assessments (National Institutes of Health Stroke Scale (NIHSS), mRS), antiplatelet medication, and quality-of-life assessment (EQ-5D-3L)). Hunt and Hess scores were recorded for patients who had evidence of aneurysm rupture and subarachnoid hemorrhage (SAH). Digital subtraction angiography (DSA) was performed at 12 months to evaluate the grade of aneurysm occlusion and parent vessel stenosis.
Primary efficacy outcome
The primary efficacy endpoint was complete aneurysm occlusion (defined as 100% occlusion of the aneurysm or Raymond-Roy (RR) class I)14 15 at 12-month follow-up DSA, in the absence of re-treatment or parent artery stenosis (>50%). Angiographic occlusion was assessed by an independent imaging core laboratory, blinded to assessments made by the clinical sites, to avoid bias. The University of California San Francisco (UCSF) interventional radiology core laboratory provided angiographic evaluation of anonymized patients in the ATLAS trial. For this specific project, the angiogram reader was SWH, who has over 15 years' experience in the interpretation of cerebral angiograms and the endovascular treatment of brain aneurysms.
Primary safety outcome
The primary safety endpoint was any incidence of major ipsilateral stroke, defined as an ipsilateral stroke with an increase of four or more points on the NIHSS assessment at 24 hours after symptoms' onset, or neurological death within 12 months postprocedure. An independent CEC adjudicated prespecified primary endpoint events and serious device-related events. The threshold for the primary safety endpoint event rate was set at <25% for the posterior circulation cohort in the ATLAS trial, which was established after reviewing data extracted from the MAPS trial3 wide-necked aneurysm patient cohort as well as published rates of procedural and long-term morbidity and mortality.
Secondary outcomes
Secondary efficacy endpoints were assessed after the index procedure through 12 months, and included procedural technical success (defined as the proportion of patients in whom the Atlas stent was successfully delivered and deployed at the target location), rates of target aneurysm occlusion across RR classes, re-treatment, recanalization, progressive occlusion of the target aneurysm, incidence of parent artery stenosis (>50% stenosis), and stent migration.
Secondary safety endpoints were any serious adverse events (SAEs) through 12 months following the procedure, including any incidence of new or worsening major ipsilateral stroke as measured by the NIHSS, device-related SAE, target aneurysm rupture, and SAH.
Statistical analysis
Descriptive statistics were compiled for baseline variables, procedural characteristics, and endpoints. Continuous and ordinal variables are summarized as mean (SD), median, and IQR. Median and interquartile range are reported when distribution of a variable is visually skewed from normal distribution. Percentages and numerators, denominators are presented for categorical and binary variables.
The proportion of patients who met the primary endpoints were compared with performance goals using the one-sided Fisher’s exact test with a significance level of α=0.025. The performance goals were determined a priori based on a meta-analysis as well as regulatory and medical considerations. Analyses for posterior circulation cohorts were performed by constructing two-sided, 95% confidence intervals about the estimates of the percentage of patients with complete aneurysm occlusion and the percentage of patients experiencing a major ipsilateral stroke or neurological death using the exact binomial (Clopper-Pearson) method. Success in the posterior circulation cohort occurred when the lower bound of the efficacy endpoint was above 50% and the upper bound of the safety endpoint was below 25%. Missing values are imputed. All analyses were performed using SAS software, version 9.4 (SAS Institute Inc., Cary, North Carolina, USA).
Results
A total of 124 patients with posterior circulation IAs were initially enrolled across 25 US centers. The modified intention-to-treat cohort included 93.5% (116/124) after the exclusion of eight patients (online supplemental figure 1). Mean patient age was 60.2±10.5 years, 81.0% (94/116) were female, and 91.4% (106/116) were Caucasian (table 1). The most frequent comorbidities were hypertension (67.2%; 78/116) and hyperlipidemia (49.1%; 57/116). The majority of patients were either current smokers (40.5%; 47/116) or past smokers (38.8%; 45/116). A history of previous hemorrhagic stroke or multiple aneurysms was reported by 17.2% (20/116) and 17.2% (20/116) of patients, respectively. A total of 113 patients (97.4%) had a baseline mRS score ≤2. Thirteen patients (11.2%) had experienced previous rupture of their target aneurysms, of whom 10 were treated with coiling only, while two were treated with balloon-assisted coiling. The other patient with a previously ruptured target aneurysm was untreated. The mean and median time from aneurysm rupture to stent placement was 800 and 189 days, respectively.
Target aneurysm characteristics are summarized in table 1.
Intraprocedural and postprocedural results
All procedures were technically successful (100.0%; 116/116). Patients were implanted with one (65.5%; 76/116) or two (34.5%; 40/116) Atlas stents. Multi-stent constructs were preplanned. Stents were implanted successfully in 94.5% (156/165) of attempts. Nine device failures occurred among five patients. All cases involved inadvertent deployment in the catheter hub (seven) or the feeling of excess friction as the stent was advanced in the microcatheter requiring removal of the catheter. None of these stents was actually deployed in a patient and all cases were successfully completed with additional devices. The median procedure duration first puncture to wound closure was 109.0 min (IQR 85.0–140.0). Immediately postprocedure, complete occlusion (RR I) was achieved in 77.6% (90/116) of patients, residual aneurysm neck filling (RR II) in 19.0% (22/116), and residual aneurysm dome filling (RR III) in 3.4% (4/116).
Primary endpoints
Of the 116 patients who completed the 12-month follow-up, 95 patients had DSA results available (table 2). The primary efficacy endpoint was achieved in 76.7% (95% CI 67.0% to 86.5%) of patients (p<0.001 vs performance goal, missing values handled with multiple imputation). The rate of complete occlusion, according to the imaging core laboratory, was 85.3% (81/95, 95% CI 76.5% to 91.7%), while parent artery stenosis >50% occurred in 2.1% (2/95, 95% CI 0.3% to 7.4%) of patients. The rate of re-treatment was 7.8% (9/116, 95% CI 3.6% to 14.2%). Of the nine patients who underwent re-treatment, 2.6% (3/116) had complete occlusion postprocedure but recanalized, 0.9% (1/116) had preplanned staged procedures (achieved complete occlusion after stage 2 operation), 2.6% (3/116) had residual neck, and 1.7% (2/116) had residual aneurysm that persisted on follow-up imaging.
The incidence of the primary safety endpoint (major ipsilateral stroke and/or neurological death) for all 116 patients was 4.3% (5/116, 95% CI 1.4% to 9.8%, p<0.001; table 2). Major ipsilateral stroke occurred in 3.4% (4/116, 95% CI 0.9% to 8.6%) of patients, while neurological death occurred in 0.9% (1/116, 95% CI 0.0% to 4.7%) of patients. This neurological death was the result of subdural hematoma and related severe pneumonia aspiration after experiencing a fall 75 days postprocedure. Of the five patients who experienced unfavorable primary safety outcomes, 1.7% (2/116) recovered with no residual deficit on subsequent follow-up,
Secondary efficacy endpoints
Secondary endpoints were evaluated at the 12-month follow-up DSA (table 3). The majority of the 95 patients with available DSA results had RR I occlusion of the target aneurysm at a rate of 85.3% (81/95, 95% CI 76.5% to 91.7%). The rate of RR II occlusion of the target aneurysm was 9.5% (9/95, 95% CI 4.4% to 17.2%), and for RR III was 5.3% (5/95, 95% CI 1.7% to 11.9%). Ninety of the 95 subjects had combined RR I and II (94.7%, 95% CI 88.1% to 98.3%).
Most patients had the same (71.6%; 68/95, 95% CI 61.4% to 80.4%) or improved (17.9%; 17/95, 95% CI 10.8% to 27.1%) occlusion status of their target aneurysms compared with immediate postprocedure RR scores. Only 10.5% (10/95, 95% CI 5.2% to 18.5%) of patients had worse occlusion status compared with immediate postprocedure RR scores. For clinical outcome at 12 months' follow-up, 93.1% (95/102) of patients had an mRS score of 0–1, while 96.1% (98/102) had an mRS score of 0–2. There was no reported occurrence of stent migration (0.0%; 0/95, 95% CI 0.0% to 3.8%).
Secondary safety endpoints
According to CEC adjudication, four patients experienced new or worsening major ipsilateral stroke (table 4), and these four patients were the same subjects who experienced the primary safety endpoint of major ipsilateral stroke. Two patients experienced SAH: one patient experienced only SAH, while the other experienced both SAH and aneurysm rupture, although neither were related to stent placement. The patient who had both SAH and aneurysm rupture also had major ipsilateral stroke and was one of the patients who experienced the primary safety endpoint of major ipsilateral stroke. Clinically, the two patients who experienced either SAH alone or both SAH and aneurysm rupture returned for their 12-month visit and their mRS score was 0. All SAEs that were site-reported as ‘possibly’, ‘probably’, or ‘related’ to the study device were classified as ‘site-reported device-related SAEs’. This included 12 SAEs in 11 subjects. For two subjects, the events were also CEC-adjudicated as primary safety endpoint events (ie, major ipsilateral stroke). Four of the 11 subjects experienced site-reported device-related SAEs that were CEC-adjudicated as minor ischemic strokes. All four subjects had excellent long-term outcomes with mRS scores of 0 or 1 and NIHSS scores of 0 at their 6- and 12-month follow-up visits. Five of the 11 subjects experienced site-reported device-related SAEs that were CEC-adjudicated as not strokes (of any type), SAH, or aneurysm rupture. Three of these patients had target aneurysm re-treatment based on asymptomatic recanalization detected on routine follow-up imaging. Two of the five subjects had transient ischemic attacks, which occurred on postoperative days 45 and 113, respectively.
Discussion
The ATLAS IDE study included a cohort of 116 patients with wide-necked IAs located in the posterior circulation, making it the largest study using Atlas stents for this particular cohort of patients. In this study, SAC using Atlas was associated with excellent technical success and safety profile rates. Atlas SAC embolization provided a high rate of 12-month complete occlusion without stenosis or re-treatment in the majority of patients, with low incidence of safety endpoints. Therefore, Atlas SAC embolization successfully met both the efficacy and safety endpoints of the trial, indicating that Atlas represents an efficacious and safe treatment option for SAC embolization of wide-necked IAs in the posterior cerebral circulation. Based on these results, the FDA awarded specific approval for use of Atlas stents in the posterior circulation on July 30, 2020, making it the only adjunctive stent in the USA that has shown safety and effectiveness in the posterior neurovasculature.
Several meta-analyses of patients with wide-necked aneurysms who underwent SAC treatment in the anterior or posterior circulation have been published.17–19 The results of these studies suggest that stenting wide-necked aneurysms with earlier generation stents can result in RR I occlusion rates in the range of 69–73%, with mortality rates less than 2.5%. Recanalization rates ranged from 9% to 13%, and one analysis reported an aggregate re-treatment rate of 5.7%.19 The results of our study demonstrate that treatment of wide-necked aneurysms with the Atlas stent results in comparable or improved outcomes compared with earlier generation stents.
Few studies have assessed efficacy and safety outcomes specific to the Atlas stent in wide-necked IAs. A recent retrospective analysis of a prospectively maintained database of 113 patients with wide-necked IAs in the anterior and posterior circulation reported a technical success rate of 100%.20 Complete occlusion (RR I) was achieved in 88% and 82% of patients immediately postprocedure and at the 12-month follow-up, respectively. At the 6-month follow-up, 96.5% (109/113) of patients had an mRS score of 0–1, with a mortality rate of 2.7% (3/113) mostly due to SAH, and a morbidity rate of 0.85%.20 Another retrospective study of 37 patients at three centers using the Atlas stent also demonstrated a technical success rate of 100%. Complete occlusion (RR I was achieved in 83.8% (31/37) and 80.8% (21/26) immediately postprocedure and at the 6-month follow-up, respectively, while neurological morbidity was reported in 2.7% (1/37) of patients at the 6-month follow-up.13 Tsai et al reported their 1-year results using Atlas stent-assisted coil embolization, which examined the efficacy and safety outcomes at discharge of 58 patients with 76 Atlas stents. In that retrospective study, 40 patients were treated with a single stent, 15 with a Y-stent, and three with an X-stent configuration.21 The immediate RR I occlusion rate was 70.7% (41/58), with no neurological morbidity and an mRS score of 0 (IQR 0–1).21 These studies were retrospective in nature, used low to moderate numbers of patients,13 21 22 did not collect long-term follow-up data,21 and included heterogeneous IAs with respect to status, size, and location. The ATLAS IDE of wide-necked IAs in the posterior circulation has examined many of these limitations, and the results are specific to the posterior circulation.
The Low-profile Visualized Intraluminal Support (LVIS and LVIS Jr) stents (MicroVention, Aliso Viejo, California, USA) were the first neurovascular stents approved for SAC embolization of anterior circulation IAs in the USA. The efficacy and safety results for the Atlas stent were similar to those of the US LVIS pivotal multicenter trial of 153 patients with wide-necked IAs in the anterior and posterior circulation at 12 months.23 Direct comparisons are difficult, but a recent study compared Atlas (37 patients) and LVIS Jr (27 patients) stent-assisted aneurysm coiling, after controlling for location, size, coiling technique, and coil packing density.12 That study observed greater rates of complete occlusion (RR I) with Atlas than with LVIS Jr (57% vs 41%), and this difference remained at follow-up (12.7 months; 100% vs 81% (17/21)). The study also detected a significantly lower rate of in-stent restenosis with Atlas (0% vs 19% (4/21)) along with similar re-treatment rates (0%).12
Limitations
The main limitation of ATLAS IDE is the single-arm nature of the study; thus, there was no direct comparative analysis with another aneurysm treatment device. However, this study had a relatively large sample size, and the study characteristics, including the independent core laboratory, CEC, and prospective study design, were all meant to mitigate potential bias.
Summary
In the ATLAS IDE posterior circulation study, Atlas SAC embolization demonstrated excellent technical success as well as promising efficacy and safety profiles in patients with wide-necked IAs in the posterior circulation. This was shown by the high rates of angiographic complete occlusion of IAs at 12 months and the low rates of serious neurological events and mortality. Therefore, the Neuroform Atlas Stent System is an effective and safe treatment option for patients with wide-necked, posterior circulation IAs.
Data availability statement
Data are available upon reasonable request. For inquiries, please contact: Brian T. Jankowitz, Associate Professor Neurosurgery Cooper University Hospital Camden, NJ 08103. Email: jankowitz-brian@CooperHealth.edu
Ethics statements
Patient consent for publication
Ethics approval
The study was performed in compliance with the World Medical Association's Declaration of Helsinki. The study protocol was approved by the local ethics committee or institutional review board.
Acknowledgments
The authors acknowledge Superior Medical Experts for editing and technical writing assistance.
References
Supplementary materials
Supplementary Data
This web only file has been produced by the BMJ Publishing Group from an electronic file supplied by the author(s) and has not been edited for content.
Footnotes
Twitter @drnimajax, @donfreimd, @GaborTothMD, @cure4stroke, @AdamArthurMD, @PeterKa80460001
Contributors All authors made substantial contributions to the conception and design, analysis, and interpretation of data; drafted or critically revised the article; and gave final approval of the version to be published.
Funding The ATLAS trial (Safety and Effectiveness of the Treatment of Wide Necked, Saccular Intracranial Aneurysms with the Neuroform Atlas Stent System) was funded by Stryker Neurovascular.
Competing interests BTJ: consultant for Stryker and a consultant/proctor for Medtronic. OOZ: research grant from Stryker, Medtronic, Cerenovus, Penumbra, and Genentech; consultant and speaker for Cerenovus, Stryker, Penumbra, and Medtronic; has ownership interest in Galaxy Therapeutics LLC; and serves on the endovascular committee as co-chair for National Institutes of Health StrokeNet Consortium and has served as an expert witness. SWH: Core Laboratory Services for Stryker and MicroVention Terumo (money paid to UCSF, over $5k); consulting for Microvention Terumo (money paid to SWH, under $5k). APJ: reports no conflicts. BG: consultant for Medtronic and Microvention. TGJ: consultant at Cerenovus (steering committee/DSMB–modest), Stryker Neurovascular (PI DAWN (DWI or CTP Assessment with Clinical Mismatch in the Triage of Wake-Up and Late Presenting Strokes Undergoing Neurointervention With Trevo)-unpaid); holds Stock at Anaconda, Blockade Medical, Route 92, Corindus, FreeOx Biotech, Viz.ai, Silk Road. JFF: equity - Fawkes Biotechnology, Cerelux, LLC; consultant: Medtronic, Penumbra, Stream Biomedical. RAH: research Grant: Medtronic, Stryker, Microvention, Cerenovus; consultant: Stryker, Medtronic, Cerenovus, Microvention; stockholder: Neurvana, Elum, EndoStream, Three Rivers Medical Inc, Rist, Cerebrotech, InNeuroCo; Scientific Advisor: MIVI, Elum, Three Rivers Medical Inc, Shape Medical. ES: reports a speakers’ agreement with Stryker. AA: reports no conflicts. DF: consultant/speakers bureau for Stryker; consultant/speakers bureau for Siemens, Penumbra, Philips, Genentech, Viz.AI; Scientific Advisory Board for Shape Memory Medical. AMM: consulting fees/honoraria: Microvention-Terumo Inc., Stryker Inc; major stock shareholder/equity: Cerevasc LLC; ownership/founder: Cerevasc LLC; intellectual property rights: Cerevasc LLC, Tufts Medical Center. AT: SCENT data safety monitoring board, funds paid to the institution. GT: consultant to Dynamed, Medtronic, and Microvention. DKL: consultant for Stryker Neurovascular. MJA: consultant for Stryker Neurovascular. AHS: financial Interest/investor/stock options/ownership for Amnis Therapeutics, Boston Scientific (for acquisition of Claret Medical), Buffalo Technology Partners, Inc., Cognition Medical, Endostream Medical, Ltd, Imperative Care, International Medical Distribution Partners, Q’Apel Medical, Inc, Rebound Therapeutics Corp., Rist Neurovascular, Inc, Serenity Medical, Inc, Silk Road Medical, Spinnaker Medical, Inc, Synchron; consultant/advisory board for Amnis Therapeutics, Boston Scientific, Cerenovus, Endostream Medical, Ltd, Imperative Care, Integra LifeSciences Corp., Medtronic, MicroVention, Penumbra, Q’Apel Medical, Inc., Rapid Medical, Rebound Therapeutics Corp, Serenity Medical, Inc, Silk Road Medical, Stryker, W.L. Gore & Associates National PI/Steering Committees: Cerenovus LARGE Trial and ARISE II Trial, mal Medtronic SWIFT PRIME and SWIFT DIRECT Trials; MicroVention FRED Trial and CONFIDENCE Study; MUSC POSITIVE Trial; Penumbra 3D Separator Trial, COMPASS Trial, INVEST Trial. PK: consultant for Stryker and Cerenovus; stockholder of InNeuroCo. MTF: grant funding from NIH, Medtronic, Microvention, Stryker, and Genentech; consultant for Medtronic, Stryker, Balt USA, Viz.ai, Genentech, Cerenovus, Corindus, and EndoPhys. AMS: research support from Stryker, Medtronic, and Penumbra; consulting for Stryker, Terumo, Penumbra, and Siemens. ASA: consultant for Johnson and Johnson, Medtronic, Microvention, Penumbra, Scientia, Siemens, Stryker; research support from Balt, Cerenovus, Medtronic, Microvention, Penumbra, Siemens, and Stryker; shareholder in Bendit, Cerebrotech, Endostream, Magneto, Marblehead, Mentice, Neurogami, Serenity, Synchron, Triad Medical. RWC: proctor for Medtronic and Cerenovus. SRS: consultant for Balt, Cerenovus, Cordis, Medtronic, Microvention, Terumo, Stryker. GPC: consultant and proctor for Stryker, MicroVention, and Medtronic. JMC: reports no disclosures. AAA: reports no disclosures. JRC: reports no disclosures. AK: reports no disclosures. RCC: consulting for Stryker. EL: reports no disclosures. AST: reports no disclosures. DL: reports no disclosures. DK: reports no disclosures. DKH: reports no disclosures. ASP: reports no disclosures. No CEC/DSMB members are from participating sites of the study. UCSF is not a participating site. RB and RK delcare none.
Provenance and peer review Not commissioned; externally peer reviewed.