Article Text

Download PDFPDF

P023 Cyclic aspiration in mechanical thrombectomy: influencing factors and experimental validation
Free
  1. Magda Jablonska1,2,
  2. Jiahui Li3,
  3. Riccardo Tiberi4,
  4. Alejandro Tomasello5,
  5. Marc Ribo1
  1. 1Stroke Unit, Vall d’Hebron University Hospital, Barcelona, Spain, Barcelona, Spain
  2. 22-nd Department of Radiology, Medical University of Gdansk, Gdansk, Poland
  3. 3Stroke Unit, Vall d’Hebron University Hospital, Barcelona, Spain, Stroke Research, Vall d’Hebron Research Institute, Barcelona, Spain, Barcelona, Spain
  4. 4Stroke Research, Vall d’Hebron Research Institute, Barcelona, Spain, Stroke Unit, Vall d’Hebron University Hospital, Barcelona, Spain, Barcelona, Spain
  5. 5Neurointerventional Radiology Department, Vall d’Hebron University Hospital, Barcelona, Spain, Barcelona, Spain

Abstract

Introduction Mechanical thrombectomy is a fundamental intervention for acute ischemic stroke treatment. While conventional techniques are effective, cyclic aspiration (CyA) shows potential for better recanalization rates.

Aim of Study We aim to investigate factors affecting CyA and compare them with static aspiration (StA).

Methods StA setup consisted of an aspiration pump connected to pressure transducer. CyA was tested with five setups: single solenoid valve with air+saline (1) or saline alone (2) as aspiration medium; two solenoid valves with air+saline (3) as aspiration medium; complete air removal and saline feeding (4); pressurized saline feeding (5). To assess the efficacy of clot ingestion, the pressure transducer was replaced with a distal aspiration catheter. Moderately stiff clot analogs (15 mm) were used to investigate the ingestion quantified as clot relative weight loss. Additionally, the aspiration flow rate was assessed for each setup.

Results With CyA setup 1, the amplitude of the achieved negative pressure waves declined with increasing frequencies but progressively increased with each subsequent iteration, achieving a maximum amplitude of 81 kPa for setup 5 at 1Hz. Relative clot weight loss was significantly higher with setup 5 at 5Hz than with StA (100% vs. 37,8%; p=0.05). Aspiration flow rate was lower with CyA than with StA (setup 5 at 5Hz: 199,8ml/min vs. StA: 311ml/min; p<0.01).

Conclusion Cyclic aspiration with the appropriate setup may represent an encouraging innovation in mechanical thrombectomy, offering a promising pathway for improving efficacy in clot ingestion and recanalization. The observed benefits warrant confirmation in a clinical setting.

Disclosure of Interest no.

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.