Article Text

Download PDFPDF
Original research
Assessment of intra-aneurysmal flow modification after flow diverter stent placement with four-dimensional flow MRI: a feasibility study
  1. Vitor Mendes Pereira1,2,3,
  2. Olivier Brina1,
  3. Benedicte M A Delattre4,
  4. Rafik Ouared1,
  5. Pierre Bouillot1,
  6. Gorislav Erceg1,
  7. Karl Schaller5,
  8. Karl-Olof Lovblad1,
  9. Maria-Isabel Vargas1
  1. 1Division of Neuroradiology, Department of Medical Imaging, University Hospitals of Geneva, Geneva, Switzerland
  2. 2Division of Neuroradiology, Department of Medical Imaging, Toronto Western Hospital, University Health Network, Toronto, Ontario, Canada
  3. 3Division of Neurosurgery, Department of Surgery, Toronto Western Hospital, University Health Network, Toronto, Ontario, Canada
  4. 4Division of Radiology, Department of Medical Imaging, University Hospitals of Geneva, Geneva, Switzerland
  5. 5Division of Neurosurgery, University Hospitals of Geneva, Geneva, Switzerland
  1. Correspondence to Dr Vitor Mendes Pereira, Division of Neuroradiology, Department of Medical Imaging and Division of Neurosurgery, Department of Surgery, Toronto Western Hospital, University Health Network, Toronto, Ontario, Canada; vitormpbr{at}hotmail.com

Abstract

Background Flow diverter stents (FDS) have been effectively used for the endovascular treatment of sidewall intracranial aneurysms (IAs). Unlike standard endovascular treatments used to exclude directly the aneurysm bulge from the parent vessel, FDS induce reduction in the intra-aneurysmal flow and promote progressive and stable thrombosis therein. The advent of FDS has therefore increased the need for understanding of IA hemodynamics.

Methods We proposed the use of the most recently evolved four-dimensional (4D) flow MRI technique to evaluate qualitatively and quantitatively post-FDS flow modification in 10 patients. We report intra-aneurysmal velocity measurements and the influence of metal artifacts induced by the stent.

Results An index was defined to quantitatively measure flow changes—namely, the proportional velocity reduction ratio (PVRR)—with ranges from 34.6% to 71.1%. Furthermore, we could compare streamlines characterizing the post-stent flow patterns in five patients in whom the intra-aneurysmal velocity was beyond the visualization threshold of 7.69 cm/s.

Conclusions Despite metal artifacts and the low velocities involved, 4D flow MRI could be of interest to measure qualitatively and quantitatively flow changes in stented aneurysms. However, further enhancements are required together with further validation work before it can be considered for clinical use.

  • Aneurysm
  • Blood Flow
  • Flow Diverter
  • MRI
  • Stent

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.