Fluoroscopic exposure during percutaneous balloon compression of the Gasserian ganglion
  1. Patrick Fransen
  1. Correspondence to Dr Patrick Fransen, Chairman, Department of Neurosurgery, Clinique du Parc Léopold, 38 rue Froissart, 1040 Brussels, Belgium; p.fransen{at}neurobrussels.be

Abstract

Aim This prospective case–control study documents the level of radiation expected during percutaneous balloon compression of the Gasserian ganglion.

Methods Percutaneous balloon compression of the Gasserian ganglion is a minimally invasive treatment for trigeminal neuralgia resistant to medical treatment and an alternative to microvascular decompression. This operation requires the use of fluoroscopy to navigate the needle through the foramen ovale and to monitor balloon inflation and deflation. However, the level of radiation exposure for this technique has never been reported. Exposure data (fluoroscopy time in seconds and radiation dose in mGycm2) were collected during 17 procedures on 14 patients using the dose report provided by the digital interface of the fluoroscope.

Results The mean fluoroscopy time per patient was 62 s, ranging from 102.9 s to 44.9 s, and the mean radiation dose per patient was 1137.18 mGycm2, ranging from 1738 mGycm2 to 639.5 mGycm2.

Conclusion This quantification of the dose and duration of radiation during percutaneous compression of the trigeminal ganglion shows that the amount of radiation required to perform this operation is significant and makes lead-coated protection clothing and careful use of the fluoroscope mandatory.

  • Trigeminal neuralgia
  • facial pain
  • percutaneous balloon compression of the gasserian ganglion
  • peroperative imaging
  • cranial nerve
  • balloon
  • selected

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Competing interests None.

  • Ethics approval Ethics approval was not required as this is a retrospective observational study not involving patient participation.

  • Provenance and peer review Not commissioned; externally peer reviewed.