Low-profile Visualized Intraluminal Support device (LVIS Jr) as a novel tool in the treatment of wide-necked intracranial aneurysms: initial experience in 32 cases
  1. Daniel Behme1,
  2. Anushe Weber1,
  3. Annika Kowoll1,
  4. Ansgar Berlis2,
  5. Thomas H Burke3,
  6. Werner Weber1,4
  1. 1Department of Radiology and Neuroradiology, Klinikum-Vest, Knappschaftskrankenhaus, Recklinghausen, Germany
  2. 2Department of Radiology and Neuroradiology, Klinikum Augsburg, Augsburg, Germany
  3. 3Microvention-Terumo, Tustin, California, USA
  4. 4Ruhr-University-Bochum, University Medical Center Knappschaftskrankenhaus Langendreer, Bochum, Germany
  1. Correspondence to Dr Anushe Weber, Department of Radiology and Neuroradiology, Klinikum Vest-Knappschaftskrankenhaus Recklinghausen, Dorstener Straße 151, Recklinghausen 45657, Germany; Anushe.Weber{at}Klinikum-Vest.de

Abstract

Background and purpose The focus of this study was to determine ease of deployment, safety and effectiveness of the LVIS Jr device.

Methods A retrospective analysis was performed of 32 cases comprising 34 aneurysms in which the LVIS Jr device was used for stent-assisted coil embolization of intracranial aneurysms from February to October 2012, including all clinical and angiographic data as well as mid-term follow-up (1–12 months of treatment).

Results The median age of the patients was 54 years (range 21–76) and 19 (59%) were women. The aneurysms were ruptured in 12/34 cases (35.3%); 26 (76.4%) were located within the anterior circulation and the remaining 8 (23.5%) were located in the posterior circulation. Eleven of the 34 aneurysms (32.3%) were treated with a Y-stent configuration. Immediate total occlusion was observed in 16/34 (47%), near total occlusion (90–95%) in 5/34 (14.7%) and a ‘dog ear’ or subtotal occlusion in 12/34 (35.2%). A single aneurysm was treated without coil embolization. Complications occurred in 5/34 cases (15%), including two cases of in-stent thrombosis.

Conclusions Implantation of the LVIS Jr device as a support device for stent-assisted coil embolization seems to be safe and effective. The LVIS Jr device can also be implanted in a Y-stent configuration, offering a novel technique with a potentially lower risk of thromboembolic complications compared with other devices.