Article Text

other Versions

Download PDFPDF
Case series
Flow diverter stents for unruptured saccular anterior circulation perforating artery aneurysms: safety, efficacy, and short-term follow-up
  1. Anna Luisa Kühn,
  2. Samuel Y Hou,
  3. Mary Perras,
  4. Christopher Brooks,
  5. Matthew J Gounis,
  6. Ajay K Wakhloo,
  7. Ajit S Puri
  1. Division of Neuroimaging and Intervention, Department of Radiology and New England Center for Stroke Research, University of Massachusetts, Worcester, Massachusetts, USA
  1. Correspondence to Dr Ajit S Puri, Division of Neuroimaging and Intervention and New England Center for Stroke Research, Department of Radiology, University of Massachusetts Medical School, 55 Lake Avenue North, Worcester, MA 01655, USA; ajit.puri{at}umassmemorial.org

Abstract

Background Anterior circulation perforating artery aneurysms including anterior choroidal artery and lenticulostriate artery aneurysms are rare. Injury to these vessels can lead to severe debilitating symptoms.

Objective To present a new approach to treatment using flow diversion technology.

Methods Patients treated with a Pipeline embolization device (PED) for perforator artery aneurysms at our institution between June 2012 and May 2013 were identified and included in our retrospective analysis. We evaluated patient vascular risk factors; family history of aneurysms; aneurysm characteristics; National Institutes of Health Stroke Scale (NIHSS) and modified Rankin Scale (mRS) on admission; and angiography follow-up and patient clinical outcome at discharge, 6 months, and 1 year.

Results We included four patients with a mean age of 59.8 years. Two patients had a positive family history of aneurysms. Patient vascular risk factors included smoking, dyslipidemia, and hypertension. All patients presented with a NIHSS and mRS of 0 on admission. Aneurysms were located at the anterior choroidal (n=2) or lenticulostriate artery (n=2) and were treated with a single PED. No periprocedural or postprocedural complications occurred. The patients were discharged with no change in NHISS or mRS score. Six-month and 1-year follow-up angiography showed complete aneurysm occlusion. Mild intimal hyperplasia was seen in 2 cases at 6 months, but was resolved at the 1-year follow-up. No re-treatment was necessary. NIHSS and mRS remained 0 at follow-up time points.

Conclusions Our preliminary results show that flow diversion technology is an effective and safe therapy for complex, hard-to-treat aneurysms in perforating arteries. Larger studies with long-term follow-up are needed to validate our promising results.

  • Aneurysm
  • Device
  • Angiography
  • Technique
  • Flow Diverter

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.