Accuracy and precision of venous pressure measurements of endovascular microcatheters in the setting of dural venous sinus stenosis
  1. Michael B Avery1,
  2. Sheryl Sambrano2,
  3. Javed Khader Eliyas1,
  4. Muneer Eesa3,
  5. Alim P Mitha1,3
  1. 1 Department of Clinical Neurosciences, University of Calgary, Calgary, Alberta, Canada
  2. 2 Department of Mechanical Engineering, University of Calgary, Calgary, Alberta, Canada
  3. 3 Department of Radiology, University of Calgary, Calgary, Alberta, Canada
  1. Correspondence to Dr Alim P Mitha, Department of Clinical Neurosciences, University of Calgary, 12th Floor Neurosciences, Foothills Medical Centre, 1403 - 29 Street N.W., Calgary, Alberta T2N 2T9, Canada; alimmitha{at}gmail.com

Abstract

Introduction Dural venous sinus stenosis (DVSS) may lead to increased intracranial pressure, sometimes requiring a stent if a high pressure gradient exists. Many neuroendovascular physicians use microcatheters to measure gradients, yet there are no studies comparing the accuracies and precisions of modern day microcatheters. We examined pressure recordings from five commonly used microcatheters in an experimental DVSS model.

Methods Using a programmable pump, dynamic flow was established in a closed circuit mimicking the venous sinus waveform. Microcatheters with 150 cm effective lengths were connected proximally to pressure transducers. Mean recording pressures were compared with a high fidelity microcatheter (HFM) in several configurations including no stenosis, proximal to a focal stenosis, and distal to a focal stenosis in opposing orientations.

Results All microcatheters recorded lower pressures than the HFM. Three of the five microcatheters successfully met intracranial pressure monitoring device standards in all conditions, while one did not meet standards in any configuration. The performance of the final microcatheter was variable, with inaccuracies occurring in unrestricted flow. All microcatheters demonstrated relatively high precision, but with variable accuracies. The larger diameter microcatheters displayed the least damping and therefore the greatest accuracies. Of the three smaller microcatheters, dimensions did not predict performance, suggesting that microcatheter construction may also play a role in pressure accuracy.

Conclusion The use of microcatheters to record dural venous sinus pressures must be done with an understanding of the inherent limitations and inaccuracies, especially if clinical decisions are made from the results.

  • angiography
  • blood pressure
  • cathether
  • intracranial pressure
  • vein

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Contributors MBA planned and conducted the statistical analysis of the data, drafted and revised the manuscript. SS contributed to the overall research design and conducted the data acquisition. JKE contributed to the research design and assisted with data acquisition. ME contributed to the research design, monitored data collection and processing, and revised the manuscript. APM contributed to the research design, monitored data collection and processing, assisted with statistical analysis of the data and revised the manuscript.

  • Competing interests APM has received grant funding from Stryker Corporation, outside of the submitted work.

  • Provenance and peer review Not commissioned; externally peer reviewed.