Abstract
Background Flow diverting stents (FDS) have transformed the treatment of intracranial aneurysms; however, their metallic structure associated with their intra-luminal positioning hamper angiographic and clinical outcomes. Therefore, there is a need to develop FDS with optimized surfaces that reduce thrombogenicity while promoting the healing process and endothelialization.
Methods P8RI, a peptide mimicking the CD31 protein, was previously developed and grafted onto Silk Vista (SV) FDS. P8RI-SV and bare-SV were used in vitro in a blood loop model to test their hemocompatibility using human whole blood and in vivo using the rabbit elastase model for optical coherence tomography (OCT) comparisons of neointimal formation at day 5 and day 28.
Results After blood loop incubation, P8RI-SV showed significant reduction in fibrin binding (p=0.004) and platelet adhesion (p=0.041) compared with bare-SV. Similarly, derivative markers measured in blood, thromboxane B2 (platelet activation) and Thrombin-Antithrombin III complexes (coagulation activation), were also significantly reduced in the P8RI-SV group (both p=0.002). In vivo, complete or near-complete occlusion was reached in all aneurysms (n=6) at day 28. Excellent rate of stent-coverage ratio was obtained at day 5 (89.3% (79.1%–98.7%)) comparable to the observation at day 28 (91.8% (79.1%–100%); p=0.44). These rates were significantly higher compared with bare-SV at day 5 (77.8% (58.3%–86.8%); p<0.001) and at day 28 (67.7% (52.6%–88.9%); p<0.0001).
Conclusion In vitro results confirm enhanced hemocompatibility with a significant anti-thrombotic effect of the P8RI-SV. In vivo results provide evidence of rapid neo-intimal growth reaching near-complete tissue healing as early as day 5 in a rabbit model.
- Aneurysm
- Flow Diverter
- Platelets
Data availability statement
Data may be obtained from a third party and are not publicly available.