TY - JOUR T1 - Safety of retained microcatheters: an evaluation of radiofrequency heating in endovascular microcatheters with nitinol, tungsten, and polyetheretherketone braiding at 1.5 T and 3 T JF - Journal of NeuroInterventional Surgery JO - J NeuroIntervent Surg SP - 314 LP - 319 DO - 10.1136/neurintsurg-2013-010746 VL - 6 IS - 4 AU - Aaron D Losey AU - Prasheel Lillaney AU - Alastair J Martin AU - Van V Halbach AU - Daniel L Cooke AU - Christopher F Dowd AU - Randall T Higashida AU - David A Saloner AU - Mark W Wilson AU - Maythem Saeed AU - Steven W Hetts Y1 - 2014/05/01 UR - http://jnis.bmj.com/content/6/4/314.abstract N2 - Background The use of ethylene-vinyl alcohol copolymer for liquid embolization of cranial vascular lesions has resulted in microcatheter fragments entrapped in patients following endovascular procedures. Undergoing subsequent diagnostic MRI examinations poses a safety concern due to the possibility of radiofrequency heating of the metallic braid incorporated into the microcatheter. Heating of nitinol, tungsten, and polyetheretherketone (PEEK) braided microcatheters was assessed and compared using a phantom model. Methods Microcatheters coupled with fluoroptic temperature probes were embedded in a polyacrylamide gel within a head and torso phantom. Experiments were performed at 1.5 T and 3 T, analyzing the effects of different catheter immersion lengths, specific absorption rate (SAR) levels, short clinical scans, long clinical scans, and microcatheter fragment lengths. Results The maximal increase in temperature for the nitinol braided microcatheter during a 15 min scan was 3.06°C using the T1 fast spin echo sequence at 1.5 T and 0.45°C using the balanced steady state free precession sequence at 3 T. The same scans for fragment lengths of 9, 18, 36, and 72 cm produced maximal temperature rises of 0.68, 0.80, 1.70, and 1.07°C at 1.5 T, respectively. The temperature changes at 3 T for these fragment lengths were 0.66, 0.83, 1.07, and 0.72°C, respectively. The tungsten and PEEK braided microcatheters did not demonstrate heating. Conclusions Substantial heating of nitinol braided microcatheters occurred and was a function of SAR level and geometric considerations. SAR and time limitations on MR scanning are proposed for patients with this microcatheter entrapped in their vasculature. In contrast, tungsten and PEEK braided microcatheters showed potential safe use in MRI. ER -