TY - JOUR T1 - AngioSuite: an accurate method to calculate aneurysm volumes and packing densities JF - Journal of NeuroInterventional Surgery JO - J NeuroIntervent Surg SP - iii28 LP - iii32 DO - 10.1136/neurintsurg-2012-010317 VL - 5 IS - Suppl 3 AU - Keith Woodward AU - David A Forsberg Y1 - 2013/11/01 UR - http://jnis.bmj.com/content/5/Suppl_3/iii28.abstract N2 - Objective AngioSuite was developed to allow the easy and accurate calculation of cerebral aneurysm volumes and packing densities. The present study was undertaken to validate its accuracy. Methods Mathematical models and software were created to facilitate more complex aneurysm volume calculations than previously described three-dimensional geometric models. 20 phantom aneurysm clay models were created and analyzed (length, width, depth and volume) using angiographic mathematical modeling (AngioSuite), geometric modeling (AngioCalc) and three-dimensional volume reconstructions from the vendor's Leonardo workstation (Siemens). Results The mean volumes of all phantoms analyzed by the AngioSuite system differed from true volumes by an average of 2.2%. Mean volumes of all phantoms analyzed by the AngioCalc system differed from true volumes by an average of 11.3%. Mean volumes obtained from the Leonardo workstation windowed for best appearance differed by an average of 5.18%. Conclusions The AngioSuite software system provides an accurate means of calculating aneurysm volumes. Typical packing densities (40% or less) are thus accurate within ±1%. Angiographic mathematical modeling with AngioSuite is more accurate than geometric modeling and three-dimensional volume reconstructions windowed for best appearance (Leonardo). ER -