RT Journal Article SR Electronic T1 Republished: Novel balloon application for rescue and realignment of a proximal end migrated pipeline flex embolization device into the aneurysmal sac: complication management JF Journal of NeuroInterventional Surgery JO J NeuroIntervent Surg FD BMJ Publishing Group Ltd. SP e4 OP e4 DO 10.1136/neurintsurg-2016-012263.rep VO 9 IS 1 A1 Mario Martínez-Galdámez A1 Joaquin Ortega-Quintanilla A1 Antonio Hermosín A1 Eduardo Crespo-Vallejo A1 Juan José Ailagas A1 Santiago Pérez YR 2017 UL http://jnis.bmj.com/content/9/1/e4.abstract AB The pipeline embolization device (PED) has become a routine firstline option for the treatment of an increasing population of intracranial aneurysms at many neurovascular centers. Intraprocedural complications during PED deployment, or complications associated with migration of the device, are rarely reported problems. Significant mismatch in luminal diameter between the inflow vessel and the outflow vessel or excessive dragging–stretching of the PED have been related to ‘watermelon seed’ or ‘accordion’ effects, respectively, resulting in stent migration. Here we present a novel balloon technique that was successfully used to realign an in situ flow diverting stent that had prolapsed into a large aneurysm. This represents a useful salvage technique and should be considered when encountering this potential complication.