RT Journal Article SR Electronic T1 Republished: Novel balloon application for rescue and realignment of a proximal end migrated pipeline flex embolization device into the aneurysmal sac: complication management JF Journal of NeuroInterventional Surgery JO J NeuroIntervent Surg FD BMJ Publishing Group Ltd. SP e4 OP e4 DO 10.1136/neurintsurg-2016-012263.rep VO 9 IS 1 A1 Martínez-Galdámez, Mario A1 Ortega-Quintanilla, Joaquin A1 Hermosín, Antonio A1 Crespo-Vallejo, Eduardo A1 Ailagas, Juan José A1 Pérez, Santiago YR 2017 UL http://jnis.bmj.com/content/9/1/e4.abstract AB The pipeline embolization device (PED) has become a routine firstline option for the treatment of an increasing population of intracranial aneurysms at many neurovascular centers. Intraprocedural complications during PED deployment, or complications associated with migration of the device, are rarely reported problems. Significant mismatch in luminal diameter between the inflow vessel and the outflow vessel or excessive dragging–stretching of the PED have been related to ‘watermelon seed’ or ‘accordion’ effects, respectively, resulting in stent migration. Here we present a novel balloon technique that was successfully used to realign an in situ flow diverting stent that had prolapsed into a large aneurysm. This represents a useful salvage technique and should be considered when encountering this potential complication.