Table 2

Summary of dynamic mechanical analyzer test conditions and results

StentManipulation before DMA testRe-sheathing during separation testForce at break (N)Main separation point
Stent 1Solitaire AB 6×20NoneNoAbout 6Proximal adhesive
Stent 2Solitaire AB 6×20NoneYesAbout 6Proximal adhesive
Stent 3Solitaire AB 6×30Deployment and retrieval 3 times in vascular replicaYesAbout 6Proximal adhesive
Stent 4Solitaire AB 6×30Deployment and retrieval once in vascular replica with pushing forward to make excessive bending at siphonNoAbout 4Distal adhesive (proximal marker at the end of wire)
Stent 5Solitaire AB 4×20Manually pulled apartNoNAThe junction point of the ball and socket joint
  • DMA, dynamic mechanical analyzer; NA, not assessable.