Table 3

Technical aspects of flow diverter devices

p64PipelineSilkFREDSurpass
Metallic component64 Braided wires stent: nitinol48 Braided wires stent: cobalt–chromium48 Braided wires stent: nitinol48 Braided wires stent: nitinol48-72-96 Braided wires stent: cobalt–chromium
Metal coverage accross the neck3.50 mm35.26%*35%35%3 mm35–36%*30%
3.75 mm37.19%*3.5 mm31–32%*
4.0 mm39.72%*4 mm29–30%*
RetrievabilityTotal (100%)Partial 75–80%Partial 75–90%Partial 75–80%Partial 75–80%
Radial forceExcellentExcellentLowExcellentGood
Radio-opacityGoodExcellentGoodGoodLow
Cost (€)10000–1200010000–1200010000–1200010000–1200010000–12000
  • *Data not available in literature. Provided by Phenox and Microvention for different device size.