Table 1

High-resolution MRI sequences and parameters at 3.0T

Parameters2D T2W2D T1W3D PD VISTA3D T2 VISTA
TE (ms)72103890
TR (ms)3550100020003000
FOV (cm)18×1818×15.818×16.525×19
Matrix448×448448×448448×413500×380
Slice thickness (mm)120.40.5
In-plane resolution (mm)0.4×0.40.4×0.350.4×0.40.5×0.5
NA3412
Acquisition time per slice (s)10.44560
Oversampling factor1.24
No of slices264–1090
Band width223207361.71.2
TSE factor22186067
Start-up echoes4164
Typical sequence time4:30 (26 slices)7:30 (10 slices)8:308:51
  • FOV, field of view; PD, proton density; TE, echo time; TR, repetition time; TSE, turbo spin echo; T1W/T2W, T1-/T2-weighted.