Table 4

Reduction in wall shear stress (WSS)

NameAve WSS
(Pa)
Max WSS
(Pa)
Reduction
(Ave WSS) (%)
Reduction
(Max WSS) (%)
Without stent3.6135.88
Pore1-d300.4412.6387.764.8
Pore1-d400.3310.8790.969.7
Pore1-d500.2610.0292.872.1
Pore2-d350.5414.2085.060.4
Pore3-d300.8714.2475.960.3
Pore3-d400.6714.2381.560.4
Pore3-d500.5313.2685.463.0
Pore4-d450.8314.3677.060.0
Pore5-d301.4912.0758.766.4
Pore5-d401.2312.4466.065.3
Pore5-d501.0413.0671.263.6
  • Ave WSS and Max WSS indicate area-averaged wall shear stress and maximum wall shear stress, respectively, on the intracranial aneurysm surface.