Table 1

Baseline characteristics of the aneurysm, parent vessel dimensions, platelet function, and the devices deployed

cPEDsPEDp Value
No DAPTDAPTNo DAPTDAPT
Aneurysm neck size (mm)4.1±1.33.8±0.73.9±1.04.3±1.80.78
Aneurysm width (mm)3.8±0.93.3±0.83.8±0.93.4±1.00.51
Aneurysm height (mm)7.7±2.26.3±1.27.6±2.57.7±2.30.33
Parent vessel diameter proximal to aneurysm (mm)3.7±0.34.0±0.63.6±0.43.9±0.40.18
Parent vessel diameter distal to aneurysm (mm)3.7±0.83.6±0.43.5±0.53.9±0.80.58
Subclavian artery diameter (mm)2.2±0.22.3±0.32.2±0.32.4±0.40.30
PRU261±4856±26249±5284±26<0.0001
ARU661±6642±43636±57657±120.90
Device diameter (mm)4.04.04.04.1250.58
Proximal landing zoneNS
  Innominate artery9999
  Herniation into aorta1111
  • All quantities expressed as mean±SD, except device diameters (median) and proximal landing zone (number of animals).

  • ARU, aspirin reaction units; cPED, classic Pipeline Embolization Device; DAPT, dual antiplatelet therapy; PRU, P2Y12 reaction units; sPED, Pipeline Flex with Shield Technology.