Phantom exposures of customized protocols
No of frames | DSA 10 s | Fluoroscopy 30 s | Roadmap 30 s | |||||||
Dose setting per frame (μGy) | DAP µGy.m2 | Estimated skin dose mGy | p/s | DAP µGy.m2 | Estimated skin dose mGy | p/s | DAP µGy.m2 | Estimated skin dose mGy | ||
Adult default vendor protocol | 30 (3 fps) | 2.4 | 730.9 | 83.6 | 15 | 45.3 | 5.2 | 15 | 101.7 | 11.6 |
Adult low dose protocol | 15 (MP) | 1.82 | 184.1 | 21.0 | 10 | 12.3 | 1.4 | 10 | 15.0 | 1.7 |
Pediatric low dose protocol | 15 (MP) | 0.54 | 36.8 | 4.2 | 7.5 | 5.8 | 0.7 | 7.5 | 15.2 | 1.7 |
Infant protocol | 15 (MP) | 0.24 | 17.9 | 2.0 | 7.5 | 5.0 | 0.6 | 7.5 | 14.5 | 1.6 |
Dose readings for dose–area product (DAP) and estimated skin dose (ESD) are listed for the four basic imaging protocols described in table 1. Readings were acquired with an adult-sized Alderson Rando head phantom in the anteroposterior projection using a 22 cm field of view setting, a source–image distance of 100 cm, collimators fully open, and with the grid in place. MP, multiphase, p/s, pulses per second.