Table 2

Phantom exposures of customized protocols

No of framesDSA
10 s
Fluoroscopy
30 s
Roadmap
30 s
Dose setting per frame
(μGy)
DAP
µGy.m2
Estimated skin dose mGyp/sDAP
µGy.m2
Estimated skin dose mGyp/sDAP
µGy.m2
Estimated skin dose mGy
Adult default vendor protocol30 (3 fps)2.4730.983.61545.35.215101.711.6
Adult low dose protocol15 (MP)1.82184.121.01012.31.41015.01.7
Pediatric low dose protocol15 (MP)0.5436.84.27.55.80.77.515.21.7
Infant protocol15 (MP)0.2417.92.07.55.00.67.514.51.6
  • Dose readings for dose–area product (DAP) and estimated skin dose (ESD) are listed for the four basic imaging protocols described in table 1. Readings were acquired with an adult-sized Alderson Rando head phantom in the anteroposterior projection using a 22 cm field of view setting, a source–image distance of 100 cm, collimators fully open, and with the grid in place. MP, multiphase, p/s, pulses per second.