Table 2

Procedural characteristics and outcome data

Overall (n=107)Catalyst 6 (n=82; 76.6%)Catalyst 6 plus additional devices (n=25; 23.4%)
General anesthesia (%)14 (13)10 (12.1)4 (16)
I.A. thrombolysis (%)2 (1.8)02 (8)
No of passes median (range)2 (1–11)2 (1–10)4 (2–11)
Procedural time median (range)40 (10–148)39 (10–110)66 (30–148)
Onset to reperfusion time median (range)272 (109–540)276 (109–540)265 (185–316)
Successful reperfusion (%)90 (84.1)74 (90.2)16 (64)
Complete reperfusion (%)68 (63.5)58 (70.7)10 (40)
Overall reperfusion (%)101 (94.3)80 (97.5)21 (84)
Unsuccessful reperfusion (%)6 (5.6)2 (2.4)4 (16)
Any intracranial hemorrhage (%)31 (28.9)22 (26.8)9 (36)
Symptomatic intracranial hemorrhage (%)4 (3.7)2 (2.4)2 (8)
24-hour clinical improvement (%)56 (52.3)51 (62.1)5 (20)
Modified Rankin Scale≤2 (%)51 (47.6)43 (52.4)8 (32)
Modified Rankin Scale≤3 (%)57 (53.2)48 (58.5)9 (36)
Mortality (%)23 (21.4)11 (13.4)12 (48)