Table 3

Diagnostic accuracy of the volumetric impedance phase shift spectroscopy device

SensitivitySpecificityAUCLR+LR−PPVNPV
Severe strokes among all strokes93% (95% CI 83 to 98)92% (95% CI 75 to 99)0.93 (95% CI 0.85 to 0.97)12.09 (95% CI 3.2 to 45.9)0.076 (95% CI 0.03 to 0.2)96% (95% CI 88 to 99)86% (95% CI 70 to 94)
Severe strokes among all evaluated subjects93% (95% CI 83 to 98)87% (95% CI 81 to 92)0.93 (95% CI 0.89 to 0.96)7.20 (95% CI 4.9 to 10.6)0.081 (95% CI 0.03 to 0.2)70% (95% CI 61 to 77)98% (95% CI 94 to 99)
  • LR +, positive likelihood ratio; LR−, negative likelihood ratio; NPV, negative predictive value; PPV, positive predictive value.