Table 2

Sensitivity and specificity of conventional and volumetric T2 in the detection of spinal vascular malformations

DiagnosisConventional T2 MRIVolumetric T2 MRI
Sensitivity, %Specificity, %PPV, %NPV, %Sensitivity, %Specificity, %PPV, %NPV, %
SVMs82.889.496.658.698.19098.190
SDAVF71.8899265.31009096.4100
SCAVM928994.58596.49096.490
  • NPV, negative predictive value; PPV, positive predictive value; SCAVM, spinal cord arteriovenous malformations; SDAVF, spinal dural arteriovenous fistula; SVMs, spinal vascular malformations.