Table 4

Efficacy and safety comparative analysis between FRED and FRED Junior devices

VariablesFREDFRED JuniorP values
n=149 (%)n=36 (%)
Clinical outcome
 mRS 3–601 (2.8)
 mRS 22 (1.3)1 (2.8)
 mRS 0–1147 (98.7)34 (94.4)0.102
Adverse events
Major complications7 (4.7)5 (13.9)
Minor complications10 (6.7)00.045
OKM grade C–D*
Total104 (87.5)28 (83.9)0.612
Follow-up >1 year81 (92%)20 (95.2)0.614
  • *In 156 cases with arterial monitoring data.