Table 2

Summary of aneurysm flow ratios between Evolve and Pipeline devices. Mean aneurysmal flow amplitude ratio values within the aneurysm sac, calculated from pre- and post Evolve and Pipeline deployment, were compared in four patient specific silicone models

ModelEvolvePipeline
MAFApreMAFApostRatioMAFApreMAFApostRatio
A2.41.60.671.91.40.74
B4.12.40.593.92.80.72
C3.42.40.712.31.70.74
D4.82.40.506.33.90.62
Average0.620.71
SD±0.09±0.06
  • MAFA, mean aneurysmal flow amplitude.