Table 3

Periprocedural complications

Flow diverter could not be delivered through microcatheter2 (2%)
Flow diverter did not open3 (3%)
Flow diverter twisted1 (1%)
Flow diverter displaced(proximally, distally)6 (6%) [4, 2]
Fish mouthing of the DED *(proximal, distal)11 (11.5%) [9, 2]
ICA dissection1 (1%)
Parent vessel occlusion(thrombus, DED related)4 (4%) [1, 3]
Multiple embolic infarcts1 (1%)
Symptomatic intracranial hemorrhage1 (1%)
Femoral artery pseudoaneurysm, groin hematoma, retroperitoneal hematoma4 (4%)
  • *Eight cases signaled by core laboratory, three cases reported by enrolling center; ICA=internal carotid artery.

  • DED, Derivo embolization device.