Use of wire as a snare for endovascular retrieval of displaced or stretched coils: rescue from a technical complication

  • Interventional Neuroradiology
  • Published:
Neuroradiology Aims and scope Submit manuscript

Abstract

Introduction

The displacement or stretching of coils during coiling of cerebral aneurysms is not an unusual technical complication, thus causing potentially disastrous consequences. Here, a simple technique using a wire as a snare for the removal of displaced and stretched coils is described.

Methods

The simple wire technique was used to remove a coil displaced in the distal M2 that a microsnare had failed to capture and another coil stretched during the coiling of cerebral aneurysms.

Results

The displaced coil lodged in the distal small cerebral vessel was entwined and removed using a microwire that was shaped manually as a pigtail at the tip. To retrieve the stretched coil, where a proximally stretched portion of the coil still remains in the delivery catheter with the distal portion placed in the aneurysm, a guidewire with a J-shaped tip was used. The coil was hooked and entwined by twisting this wire tip, which could be removed without difficulty.

Conclusion

This simple technique using a wire as a snare could be a useful method for removing displaced or stretched coils in selected cases.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2

References

  1. Fiorella D, Albuquerque FC, Deshmukh VR, McDougall CG (2005) Monorail snare technique for the recovery of stretched platinum coils: technical case report. Neurosurgery 57(1 Suppl):E210

    Article  PubMed  Google Scholar 

  2. Graves VB, Rappe AH, Smith TP, Sepetka I, Ahuja A, Strother CM (1993) An endovascular retrieving device for use in small vessels. AJNR Am J Neuroradiol 14:804–808

    CAS  PubMed  Google Scholar 

  3. Henkes H, Lowens S, Preiss H, Reinartz J, Miloslavski E, Kühne D (2006) A new device for endovascular coil retrieval from intracranial vessels: alligator retrieval device. AJNR Am J Neuroradiol 27:327–329

    CAS  PubMed  Google Scholar 

  4. Koseoglu K, Parildar M, Oran I, Memis A (2004) Retrieval of intravascular foreign bodies with goose neck snare. Eur J Radiol 49:281–285

    Article  PubMed  Google Scholar 

  5. Prestigiacomo CJ, Fidlow K, Pile-Spellman J (1999) Retrieval of a fractured Guglielmi detachable coil with use of the goose neck snare “twist” technique. J Vasc Interv Radiol 10:1243–1247

    Article  CAS  PubMed  Google Scholar 

  6. Raftopoulos C, Goffette P, Billa RF, Mathurin P (2002) Transvascular coil hooking procedure to retrieve an unraveled Guglielmi detachable coil: technical note. Neurosurgery 50:912–914

    Article  PubMed  Google Scholar 

  7. Schütz A, Solymosi L, Vince GH, Bendszus M (2005) Proximal stent fixation of fractured coils: technical note. Neuroradiology 47:874–878

    Article  PubMed  Google Scholar 

  8. Smith TP, Graves VB, Halbach VV, Higashida RT, Fraser KW, Dowd CF, Hieshima GB (1993) Microcatheter retrieval device for intravascular foreign body removal. AJNR Am J Neuroradiol 14:809–811

    CAS  PubMed  Google Scholar 

  9. Standard SC, Chavis TD, Wakhloo AK, Ahuja A, Guterman LR, Hopkins LN (1994) Retrieval of a Guglielmi detachable coil after unraveling and fracture: case report and experimental results. Neurosurgery 35:994–999

    Article  CAS  PubMed  Google Scholar 

  10. Thornton J, Dovey Z, Alazzaz A, Misra M, Aletich VA, Debrun GM, Ausman JI, Charbel FT (2000) Surgery following endovascular coiling of intracranial aneurysms. Surg Neurol 54:352–360

    Article  CAS  PubMed  Google Scholar 

  11. Vora N, Thomas A, Germanwala A, Jovin T, Horowitz M (2008) Retrieval of a displaced detachable coil and intracranial stent with an L5 Merci Retriever during endovascular embolization of an intracranial aneurysm. J Neuroimaging 18:81–84

    Article  PubMed  Google Scholar 

  12. Watanabe A, Hirano K, Mizukawa K, Kamada M, Okamura H, Suzuki Y, Ishii R (1995) Retrieval of a migrated detachable coil—case report. Neurol Med Chir (Tokyo) 35:247–250

    Article  CAS  Google Scholar 

  13. Zoarski GH, Bear HM, Clouston JC, Ragheb J (1997) Endovascular extraction of malpositioned fibered platinum microcoils from the aneurysm sac during endovascular therapy. AJNR Am J Neuroradiol 18:691–695

    CAS  PubMed  Google Scholar 

Download references

Conflict of interest statement

I declare that I have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Chang-Young Lee.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Lee, CY. Use of wire as a snare for endovascular retrieval of displaced or stretched coils: rescue from a technical complication. Neuroradiology 53, 31–35 (2011). https://doi.org/10.1007/s00234-010-0679-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00234-010-0679-4

Keywords

Navigation