Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT

  • Neuro
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract

Objective

We tested the hypothesis that Flat Detector computed tomography (FD-CT) with intravenous contrast medium would allow the calculation of whole brain cerebral blood volume (CBV) mapping (FD-CBV) and would correlate with multislice Perfusion CT (PCT).

Methods

Twenty five patients were investigated with FD-CBV and PCT. Correlation of the CBV maps of both techniques was carried out with measurements from six anatomical regions from both sides of the brain. Mean values of each region and the correlation coefficient were calculated. Bland-Altman analysis was performed to compare the two different imaging techniques.

Results

The image and data quality of both PCT and FD-CBV were suitable for evaluation in all patients. The mean CBV values of FD-CBV and PCT showed only minimal differences with overlapping standard deviation. The correlation coefficient was 0.79 (p < 0.01). Bland-Altman analysis showed a mean difference of -0.077 ± 0.48 ml/100 g between FD-CBV and PCT CBV measurements, indicating that FD-CBV values were only slightly lower than those of PCT.

Conclusion

CBV mapping with intravenous contrast medium using Flat Detector CT compared favourably with multislice PCT. The ability to assess cerebral perfusion within the angiographic suite may improve the management of ischaemic stroke and evaluation of the efficacy of dedicated therapies.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Fahrig R, Fox S, Lownie S, Holdsworth DW (1997) Use of a C-arm system to generate true 3-D computed rotational angiograms: Preliminary in vitro and in vivo results. AJNR Am J Neuroradiol 18:1507–1514

    PubMed  CAS  Google Scholar 

  2. Akpek S, Brunner T, Benndorf G, Strother C (2005) Three-dimensional imaging and cone beam volume CT in C-arm angiography with flat panel detector. Diagn Interv Radiol 11:10–13

    PubMed  Google Scholar 

  3. Kalender W, Kyriakou Y (2007) Flat-detector CT (FD-CT). Eur Radiol 17:2767–2779

    Article  PubMed  Google Scholar 

  4. Zellerhof M, Scholz B, Rührnschopf E (2005) Low contrast 3D-reconstruction from C-arm data. In Flynn M (ed) Medical Imaging 2005: Physics of Medical Imaging. Proc SPIE 5745:646-655 doi:10.1117/12.593433

  5. Doelken M, Struffert T, Richter G, Engelhorn T, Nimsky C, Ganslandt O et al (2008) Flat-panel detector volumetric CT for visualization of subarachnoid hemorrhage and ventricles: preliminary results compared to conventional CT. Neuroradiology 50:517–523

    Article  PubMed  CAS  Google Scholar 

  6. Struffert T, Richter G, Engelhorn T, Doelken M, Goelitz P, Kalender W et al (2009) Visualisation of intracerebral haemorrhage with flat-detector CT compared to multislice CT: results in 44 cases. Eur Radiol 19:619–25

    Article  PubMed  Google Scholar 

  7. Söderman M, Babic D, Holmin S, Andersson T (2008) Brain imaging with a flat detector C-arm: Technique and clinical interest of XperCT. Neuroradiology 50:863–898

    Article  PubMed  Google Scholar 

  8. Engelhorn T, Struffert T, Richter G, Doelken M, Ganslandt O, Kalender W et al (2008) Flat-panel detector angiographic CT in the management of aneurysmal rupture during coil embolization. AJNR Am J Neuroradiol 29:1581–1584

    Article  PubMed  CAS  Google Scholar 

  9. Zellerhoff M, Deuerling-Zheng Y, Strother CM, Ahmed AS, Pulfer K, Redel T, et al. (2009) Measurement of cerebral blood volume using angiographic C-arm systems. In Hu XP, Clough AV (eds) Medical Imaging 2009: Biomedical Applications in Molecular, Structural, and Functional Imaging. Proc SPIE 7262:72620H-72620H-8

  10. Hamberg LM, Hunter GJ, Kierstead D, Lo EH, Gonzalez RG, Wolf GL (1996) Measurement of cerebral blood volume with subtraction three-dimensional functional CT. AJNR Am J Neuroradiol 17:1861–1869

    PubMed  CAS  Google Scholar 

  11. Hunter GJ, Silvennoinen HM, Hamberg LM, Koroshetz WJ, Buonanno FS, Schwamm LH et al (2003) Whole-brain CT perfusion measurement of perfused cerebral blood volume in acute ischemic stroke: probability curve for regional infarction. Radiology 227:725–730

    Article  PubMed  Google Scholar 

  12. Kloska S, Fischer T, Nabavi D, Dittrich R, Ditt H, Klotz E et al (2007) Color-coded perfused blood volume imaging using multidetector CT: initial results of whole-brain perfusion analysis in acute cerebral ischemia. Eur Radiol 17:2352–2358

    Article  PubMed  Google Scholar 

  13. Ahmed AS, Zellerhoff M, Strother CM, Pulfer KA, Redel T, Deuerling-Zheng Y et al (2009) C-arm CT measurement of cerebral blood volume: an experimental study in canines. AJNR Am J Neuroradiol 30:917–922

    Article  PubMed  CAS  Google Scholar 

  14. Bley T, Strother CM, Pulfer K, Royalty K, Zellerhoff M, Deuerling-Zheng Y et al (2010) C-Arm CT measurement of cerebral blood volume in ischemic stroke: an experimental study in canines. AJNR Am J Neuroradiol 31:536–540

    Article  PubMed  CAS  Google Scholar 

  15. Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender W et al (2010) Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol. doi:10.3174/ajnr.A2083

    Google Scholar 

  16. Klotz E, Koenig M (1999) Perfusion measurements of the brain: using dynamic CT for the quantitative assessment of cerebral ischemia in acute stroke. Eur J Radiol 30:170–184

    Article  PubMed  CAS  Google Scholar 

  17. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–3010

    PubMed  CAS  Google Scholar 

  18. Schellinger PD, Fiebach JB, Hacke W (2003) Imaging-based decision making in thrombolytic therapy for ischemic stroke: present status. Stroke 34:575–583

    Article  PubMed  Google Scholar 

  19. Muir KW, Buchan A, von Kummer R, Rother J, Baron JC (2005) Imaging of acute stroke. Lancet Neurol 5:755–768

    Article  Google Scholar 

  20. Wintermark M, Flanders AE, Velthuis B, Meuli R, van Leeuwen M, Goldsher D et al (2006) Perfusion-CT assessment of infarct core and penumbra: receiver operating characteristic curve analysis in 130 patients suspected of acute hemispheric stroke. Stroke 37:979–985

    Article  PubMed  Google Scholar 

  21. Parsons MW, Pepper EM, Bateman GA, Wang Y, Levi CR (2007) Identification of the penumbra and infarct core on hyperacute noncontrast and perfusion CT. Neurology 68:730–736

    Article  PubMed  CAS  Google Scholar 

  22. Murphy BD, Fox AJ, Lee DH, Sahlas DJ, Black SE, Hogan MJ et al (2008) White matter thresholds for ischemic penumbra and infarct core in patients with acute stroke: CT perfusion study. Radiology 247:818–825

    Article  PubMed  Google Scholar 

  23. Broderick JP (2009) Endovascular therapy for acute ischemic stroke. Stroke 40(3 Suppl):S103–106

    Article  PubMed  Google Scholar 

  24. Khatri P, Abruzzo T, Yeatts SD, Nichols C, Broderick JP, Tomsick TA, IMS I, Investigators II (2009) Good clinical outcome after ischemic stroke with successful revascularization is time-dependent. Neurology 73:1066–1072

    Article  PubMed  CAS  Google Scholar 

  25. Orth R, Wallace M, Kuo M (2008) C-arm cone-beam CT: general principles and technical considerations for use in interventional radiology. J Vasc Interv Radiol 19:814–820

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Tobias Struffert.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Struffert, T., Deuerling-Zheng, Y., Kloska, S. et al. Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT. Eur Radiol 21, 882–889 (2011). https://doi.org/10.1007/s00330-010-1957-6

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00330-010-1957-6

Keyword

Navigation