Vascular complications after percutaneous coronary intervention following hemostasis with the Mynx vascular closure device versus the AngioSeal vascular closure device

J Invasive Cardiol. 2010 Apr;22(4):175-8.

Abstract

We investigated the prevalence of vascular complications after PCI following hemostasis in 190 patients (67% men and 33% women, mean age 64 years) treated with the AngioSeal vascular closure device (St. Jude Medical, Austin, Texas) versus 238 patients (67% men and 33% women, mean age 64 years) treated with the Mynx vascular closure device (AccessClosure, Mountain View, California).

Results: Death, myocardial infarction or stroke occurred in none of the 190 patients (0%) treated with the AngioSeal versus none of 238 patients (0%) treated with the Mynx. Major vascular complications occurred in 4 of 190 patients (2.1%) treated with the AngioSeal versus 5 of 238 patients (2.1%) treated with the Mynx (p not significant). Major vascular complications in patients treated with the AngioSeal included removal of a malfunctioning device (1.1%), hemorrhage requiring intervention (0.5%) and hemorrhage with a loss of > 3g Hgb (0.5%). The major vascular complications in patients treated with the Mynx included retroperitoneal bleeding requiring surgical intervention (0.8%), pseudoaneurysm with surgical repair (0.8%) and hemorrhage with a loss of > 3g Hgb (0.4%). These complications were not significantly different between the two vascular closure devices (p = 0.77). Minor complications included hematoma > 5 cm (0.5%, n = 1) within the AngioSeal group, as well as procedure failure requiring > 30 minutes of manual compression after device deployment, which occurred in 7 out of 190 patients (3.7%) treated with the AngioSeal versus 22 of 238 patients with the Mynx (9.2%) (p = 0.033).

Conclusions: Major vascular complications after PCI following hemostasis with vascular closure devices occurred in 2.1% of 190 patients treated with the AngioSeal vascular closure device versus 2.1% of 238 patients treated with the Mynx vascular closure device (p not significant). The Mynx vascular closure device appears to have a higher rate of device failure.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aneurysm, False / epidemiology*
  • Aneurysm, False / etiology*
  • Aneurysm, False / surgery
  • Angioplasty, Balloon, Coronary / adverse effects*
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Cardiac Catheterization / instrumentation*
  • Cross-Sectional Studies
  • Equipment Design
  • Equipment Failure Analysis
  • Female
  • Hemostasis, Surgical / adverse effects*
  • Hemostasis, Surgical / instrumentation*
  • Humans
  • Male
  • Middle Aged
  • Postoperative Complications / epidemiology*
  • Postoperative Complications / etiology*
  • Postoperative Complications / surgery
  • Postoperative Hemorrhage / epidemiology*
  • Postoperative Hemorrhage / etiology*
  • Postoperative Hemorrhage / surgery
  • Reoperation
  • Retroperitoneal Space / surgery