Comaneci-Assisted Coiling as a Treatment Option for Acutely Ruptured Wide Neck Cerebral Aneurysm: Case Series of 118 Patients

Neurosurgery. 2020 Nov 16;87(6):1148-1156. doi: 10.1093/neuros/nyaa200.

Abstract

Background: Wide-necked cerebral aneurysms in the setting of acute subarachnoid hemorrhage (SAH) remain difficult to treat with endovascular methods despite recent progress in the neuroendovascular field.

Objective: To evaluate the effectiveness and safety of the Comaneci device (Rapid Medical, Israel) in endovascular coil embolization of acutely ruptured, wide-necked sidewall, or bifurcation cerebral aneurysms.

Methods: We retrospectively reviewed 45 anterior communicating artery, 24 internal carotid artery, 21 middle cerebral artery bifurcation, 15 anterior cerebral artery, and 13 posterior circulation aneurysms, which were treated using Comaneci-assisted coil embolization from August 2017 to January 2019. We evaluated procedural complications, clinical outcomes, and mid-term angiographic follow-up. Immediate and 90 d-clinical outcome and radiological follow-up were obtained in all patients.

Results: Comaneci-assisted coil embolization was performed in 118 acutely ruptured aneurysms. The technique was carried out successfully in all cases. Simultaneous application of 2 separated Comaneci devices was performed in 8/118 cases (6.77%). Periprocedural thromboembolic complications related to the device were seen in 7/118 cases (5.93%) and severe vasospasm of the parent artery after manipulation of the Comaneci device occurred in 5/118 cases (4.2%). The procedural-related morbidity rate was 2.54%, and there was no procedural related mortality. Among the available survivors, angiographic follow-ups were obtained at 3 and 6 mo, and complete aneurysmal obliteration was confirmed in 81/112 (72.3%) and 75/112 (66.9%) cases, respectively. Mid-term follow-up reviewed total recanalization rate of 14.28%.

Conclusion: Comaneci-assisted embolization of wide-necked intracranial aneurysms in patients presenting with acute SAH is associated with high procedural safety and adequate occlusion rates. Furthermore, dual antiplatelet therapy can be safely avoided in this patient group.

Keywords: Bifurcation aneurysms; Coil embolization; Comaneci device; Endovascular remodeling; Wide-necked aneurysms.

MeSH terms

  • Aneurysm, Ruptured* / diagnostic imaging
  • Aneurysm, Ruptured* / therapy
  • Cerebral Angiography
  • Embolization, Therapeutic* / adverse effects
  • Humans
  • Intracranial Aneurysm* / diagnostic imaging
  • Intracranial Aneurysm* / therapy
  • Retrospective Studies
  • Stents
  • Treatment Outcome